October 2009 9650-1211-01 Rev. HPulse CO-Oximetry(SpO2, SpCO, SpMet)
E Series - Pulse CO-Oximetry (SpO2, SpCO, and SpMet) Option Insert8 9650-1211-01 Rev. HThe SpO2 Menu (SpCO/SpMet not installed)When the SpO2 parameter
Alarms9650-1211-01 Rev. H 9SpMet softkeyPressing this softkey will instantly display the SpMet reading in the CO-oximetry data field. It will displa
E Series - Pulse CO-Oximetry (SpO2, SpCO, and SpMet) Option Insert10 9650-1211-01 Rev. HDefault SettingsWhen the pulse CO-oximeter is turned on, the
CO-Oximeter Accessories9650-1211-01 Rev. H 11Red DCI-dc3 3' Adult Reusable Patient Cable / Sensor for monitoring SpO2. 8000-0332Red DCIP-dc3 3&
E Series - Pulse CO-Oximetry (SpO2, SpCO, and SpMet) Option Insert12 9650-1211-01 Rev. HMessages and TroubleshootingThe following table lists the err
Specifications9650-1211-01 Rev. H 13SpecificationsGeneralRange Oxygen Saturation (% SpO2) 1% - 100%Carboxyhemoglobin Saturation (% SpCO) 0% - 99%M
E Series - Pulse CO-Oximetry (SpO2, SpCO, and SpMet) Option Insert14 9650-1211-01 Rev. HNote: The E Series Pulse Oximetry Option is calibrated for fu
The issue date or revision level for this operation guide is shown on the front cover.ZOLL and E Series are registered trademarks of ZOLL Medical Corp
9650-1211-01 Rev. H 1PULSE CO-OXIMETRY (SpO2, SpCO, SpMet)General InformationProduct DescriptionThe E Series® pulse CO-oximeter continuously and non
E Series - Pulse CO-Oximetry (SpO2, SpCO, and SpMet) Option Insert2 9650-1211-01 Rev. HSafety ConsiderationsWarningsGeneral• Before use, carefully rea
Pulse CO-Oximetry Indications for Use9650-1211-01 Rev. H 3Pulse CO-Oximetry Indications for UseThe ZOLL E Series pulse CO-oximeter, with Masimo Rain
E Series - Pulse CO-Oximetry (SpO2, SpCO, and SpMet) Option Insert4 9650-1211-01 Rev. H4. Connect the patient cable to the CO-oximetry connector at th
Ensuring Accurate Monitoring9650-1211-01 Rev. H 5cable’s male connector all the way into the patient cable connector (see Figure 5). Figure 55. Lowe
E Series - Pulse CO-Oximetry (SpO2, SpCO, and SpMet) Option Insert6 9650-1211-01 Rev. H• Ensure that blood flow is not restricted when securing a sens
How SpO2 , SpCO and SpMet measurements are Displayed9650-1211-01 Rev. H 7How SpO2 , SpCO and SpMet measurements are DisplayedThe pulse CO-oximetry o