Page 1 - MD Operator’s Guide

Propaq® MD Operator’s Guide9650-0805-01 Rev. A

Page 2

TABLE OF CONTENTSviii www.zoll.com 9650-0805-01 Rev. AAppendix A SpecificationsDefibrillator...

Page 3 - Table of Contents

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-4 www.zoll.com 9650-0805-01- Rev. ASelecting the NIBP CuffTo take accurate measurements, you

Page 4 - Chapter 4 Trends

Selecting the NIBP Cuff9650-0805-01- Rev. A Propaq MD Operator’s Guide 8-5Caution Selection of the correct cuff is critical to the accuracy of NIBP me

Page 5 - Chapter 6 Monitoring ECG

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-6 www.zoll.com 9650-0805-01- Rev. AConnecting the NIBP CuffConnecting the NIBP cuff requires

Page 6 - Chapter 10 Monitoring CO2

Connecting the NIBP Cuff9650-0805-01- Rev. A Propaq MD Operator’s Guide 8-7Attach a single lumen hose as follows:Figure 8-2 Attaching Single Lumen Hos

Page 7

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-8 www.zoll.com 9650-0805-01- Rev. A2. Insert the plastic connectors on the NIBP hose into the

Page 8 - Chapter 15 External Pacing

Ensuring Correct Cuff Inflation Settings9650-0805-01- Rev. A Propaq MD Operator’s Guide 8-9Caution • Using a cuff that is loosely applied or too small

Page 9 - Chapter 18 Maintenance

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-10 www.zoll.com 9650-0805-01- Rev. AIn order to accurately measure systolic pressure, the CIT

Page 10 - Appendix B Accessories

Configuring NIBP Alarms and Settings9650-0805-01- Rev. A Propaq MD Operator’s Guide 8-115. On the selected NIBP alarm settings menu, use the navigatio

Page 11 - General Information

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-12 www.zoll.com 9650-0805-01- Rev. ASetting Upper and Lower NIBP MAP Alarm LimitsInitially, t

Page 12 - Unpacking

Configuring NIBP Alarms and Settings9650-0805-01- Rev. A Propaq MD Operator’s Guide 8-13On the NIBP Parameter Control Panel, you can select an alarm c

Page 13 - Symbols Used on the Equipment

9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-1Chapter 1General InformationProduct DescriptionThe ZOLL® Propaq® MD unit is an easy-to-use portable

Page 14 - CHAPTER 1GENERAL INFORMATION

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-14 www.zoll.com 9650-0805-01- Rev. ABy default, Smartcuf is Enabled.There are situations wher

Page 15

NIBP System Messages9650-0805-01- Rev. A Propaq MD Operator’s Guide 8-15NIBP System MessagesWhen monitoring NIBP, the Propaq MD unit may display the f

Page 16 - Propaq MD Indications for Use

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-16 www.zoll.com 9650-0805-01- Rev. A

Page 17 - ECG Monitoring

9650-0805-01 Rev. A Propaq MD Operator’s Guide 9-1Chapter 9Monitoring Pulse Oximetry (SpO2)The Propaq MD SpO2 input is Type CF defibrillator proof.Thi

Page 18

CHAPTER 9MONITORING PULSE OXIMETRY (SPO2)9-2 www.zoll.com 9650-0805-01 Rev. AWarnings -- SpO2, General • SpO2 style pulse oximeter measurements may b

Page 19 - Propaq MD Product Functions

SpO2 Setup and Use9650-0805-01 Rev. A Propaq MD Operator’s Guide 9-3SpO2 Setup and UseTo take accurate SpO2 measurements using the Propaq MD unit, you

Page 20 - Batteries

CHAPTER 9MONITORING PULSE OXIMETRY (SPO2)9-4 www.zoll.com 9650-0805-01 Rev. AApplying a Reusable SpO2 SensorThe following instructions describe how t

Page 21

Applying the SpO2 Sensor9650-0805-01 Rev. A Propaq MD Operator’s Guide 9-5Applying a Single-Use SensorWhen applying a single-use SpO2 sensor, be caref

Page 22 - Ready For Use (RFU) Indicator

CHAPTER 9MONITORING PULSE OXIMETRY (SPO2)9-6 www.zoll.com 9650-0805-01 Rev. AConnecting the SpO2 SensorTo connect the SpO2 sensor to the Propaq MD un

Page 23 - Warnings

Displaying SpO2 Measurements9650-0805-01 Rev. A Propaq MD Operator’s Guide 9-73. Lower the clear plastic cover over the connection to secure it:Figure

Page 24

CHAPTER 1GENERAL INFORMATION1-2 www.zoll.com 9650-0805-01 Rev. APropaq MD Optional FeaturesThe following features are optional in the Propaq MD unit.

Page 25 - Defibrillation

CHAPTER 9MONITORING PULSE OXIMETRY (SPO2)9-8 www.zoll.com 9650-0805-01 Rev. A6. When you are finished changing values on the alarm menu, navigate to

Page 26

SpO2 System Messages9650-0805-01 Rev. A Propaq MD Operator’s Guide 9-9Selecting the Heart Rate/ Pulse Rate (HR/PR) ToneThe unit allows you enable or d

Page 27

CHAPTER 9MONITORING PULSE OXIMETRY (SPO2)9-10 www.zoll.com 9650-0805-01 Rev. AFunctional Testers and Patient SimulatorsSome models of commercially av

Page 28 - Noninvasive Blood Pressure

9650-0805-01 Rev. A Propaq MD Operator’s Guide 10-1Chapter 10Monitoring CO2This chapter describes how to use the Propaq MD unit to monitor End Tidal C

Page 29 - Ferromagnetic Equipment

CHAPTER 10 MONITORING CO210-2 www.zoll.com 9650-0805-01 Rev. AWarning! • When using a sampling line for intubated patients with a closed suction syst

Page 30 - Operator Safety

CO2 Monitoring Setup and Use9650-0805-01 Rev. A Propaq MD Operator’s Guide 10-3Selecting the CO2 Sampling LineTo select the correct CO2 sampling line,

Page 31 - Patient Safety

CHAPTER 10 MONITORING CO210-4 www.zoll.com 9650-0805-01 Rev. AConnecting the CO2 Sampling LinesTo connect the FilterLine or Smart CapnoLine:1. Slide

Page 32 - Cautions

CO2 Monitoring Setup and Use9650-0805-01 Rev. A Propaq MD Operator’s Guide 10-5WAR NIN G! • The exhaust port for the Microstream sidestream CO2 appara

Page 33 - FDA Tracking Requirements

CHAPTER 10 MONITORING CO210-6 www.zoll.com 9650-0805-01 Rev. AApplying a Smart CapnoLine Nasal or Nasal/Oral CannulaThe nasal and nasal/oral cannulas

Page 34 - Software License

Measuring CO29650-0805-01 Rev. A Propaq MD Operator’s Guide 10-7Measuring CO2Once setup is complete, press to begin monitoring CO2. The numeric CO2

Page 35

Symbols Used on the Equipment9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-3Symbols Used on the EquipmentAny or all of the following symbols may be

Page 36 - The ZOLL Serial Number

CHAPTER 10 MONITORING CO210-8 www.zoll.com 9650-0805-01 Rev. ASetting CO2 and Respiration Rate AlarmsThe Propaq MD unit sounds alarms whenever measur

Page 37 - Product Overview

Setting CO2 and Respiration Rate Alarms9650-0805-01 Rev. A Propaq MD Operator’s Guide 10-9Setting Upper and Lower EtCO2 LimitsInitially, the EtCO2 Ala

Page 38 - The Front Panel

CHAPTER 10 MONITORING CO210-10 www.zoll.com 9650-0805-01 Rev. ASetting Upper and Lower Respiratory Rate (RR/BR) LimitsInitially, the RR/BR Alarm menu

Page 39

System Messages9650-0805-01 Rev. A Propaq MD Operator’s Guide 10-11Setting the CO2 Sweep SpeedThe EtCO2 sweep speed determines the X-axis scale of the

Page 40 - Display Screen

CHAPTER 10 MONITORING CO210-12 www.zoll.com 9650-0805-01 Rev. APatentsThe capnography component of the Propaq MD unit is covered by one or more of th

Page 41 - Propaq MD Display Screen

9650-0805-01 Rev. A Propaq MD Operator’s Guide 11-1Chapter 11Monitoring Invasive Pressures (IBP)The Propaq MD IBP inputs are Type CF defibrillator pro

Page 42 - CHAPTER 2PRODUCT OVERVIEW

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-2 www.zoll.com 9650-0805-01 Rev. AWarning! • If electrocautery is used, always avoid using any trans

Page 43 - Patient Cables and Connectors

Zeroing the Transducer9650-0805-01 Rev. A Propaq MD Operator’s Guide 11-3Figure 11-1 Plugging the Transducer into the Propaq MDWhen you plug the trans

Page 44 - 1. Insert MFC into unit

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-4 www.zoll.com 9650-0805-01 Rev. AFigure 11-2 IBP Channel Parameter Control Panel6. Select Zero Prob

Page 45 - 1. Align MFC as shown

Displaying IBP Measurements9650-0805-01 Rev. A Propaq MD Operator’s Guide 11-5Displaying IBP MeasurementsAfter attaching and zeroing a transducer, the

Page 46

CHAPTER 1GENERAL INFORMATION1-4 www.zoll.com 9650-0805-01 Rev. AEquipotentiality.Alternating current (ac).Direct current (dc).Auxiliary power adapter

Page 47 - Auxiliary Power Adapter

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-6 www.zoll.com 9650-0805-01 Rev. AEnabling/Disabling IBP Alarms and Setting Alarm LimitsWhen enabled

Page 48 - Navigating the Display Screen

Enabling/Disabling IBP Alarms and Setting Alarm Limits9650-0805-01 Rev. A Propaq MD Operator’s Guide 11-7Setting Upper and Lower Diastolic (DIA) Alarm

Page 49

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-8 www.zoll.com 9650-0805-01 Rev. ASetting IBP Source LabelIn the IBP channel menu, you can select a

Page 50

IBP System Messages9650-0805-01 Rev. A Propaq MD Operator’s Guide 11-9IBP System MessagesThe Propaq MD unit may display the following messages when mo

Page 51 - Display Brightness

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-10 www.zoll.com 9650-0805-01 Rev. A

Page 52 - Common Tasks

9650-0805-01 Rev. A Propaq MD Operator’s Guide 12-1Chapter 12Monitoring TemperatureThe Propaq MD Temperature inputs are Type CF defibrillator proof.Th

Page 53 - Using Treatment Buttons

CHAPTER 12 MONITORING TEMPERATURE12-2 www.zoll.com 9650-0805-01 Rev. AWarning! • The application and use of metal-jacketed temperature probes that co

Page 54

Enabling/Disabling Temperature Alarms and Setting Alarm Limits9650-0805-01 Rev. A Propaq MD Operator’s Guide 12-3Enabling/Disabling Temperature Alarms

Page 55 - Monitoring Overview

CHAPTER 12 MONITORING TEMPERATURE12-4 www.zoll.com 9650-0805-01 Rev. ASetting Upper and Lower Temperature Alarm LimitsInitially, the Temperature

Page 56 - CHAPTER 3MONITORING OVERVIEW

Temperature System Messages9650-0805-01 Rev. A Propaq MD Operator’s Guide 12-5If you don’t select a label, the Temperature channels appear with the de

Page 57 - Monitoring Display Options

Symbols Used on the Equipment9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-5Do not discard in trash. Recycle or dispose of properly.Return to a col

Page 58

CHAPTER 12 MONITORING TEMPERATURE12-6 www.zoll.com 9650-0805-01 Rev. A

Page 59

9650-0805-01 Rev. A Propaq MD Operator’s Guide 13-1Chapter 13Monitoring 12-Lead ECGThe Propaq MD 12-Lead input is Type CF defibrillator proof.This cha

Page 60

CHAPTER 13 MONITORING 12-LEAD ECG13-2 www.zoll.com 9650-0805-01 Rev. A• When not in use, cover the patient cable's V-lead connector with the sup

Page 61

Entering Patient Information9650-0805-01 Rev. A Propaq MD Operator’s Guide 13-3To enter patient information, use the navigation keys to highlight and

Page 62

CHAPTER 13 MONITORING 12-LEAD ECG13-4 www.zoll.com 9650-0805-01 Rev. AEntering Patient Age and GenderThe Patient Info parameter panel provides defaul

Page 63

12-Lead ECG Monitoring Setup9650-0805-01 Rev. A Propaq MD Operator’s Guide 13-5Applying Electrodes to the PatientDepending on local usage, ECG lead wi

Page 64

CHAPTER 13 MONITORING 12-LEAD ECG13-6 www.zoll.com 9650-0805-01 Rev. AAvoid placing electrodes over tendons and major muscle masses.Make sure that th

Page 65 - Chapter 4

12-Lead ECG Monitoring Setup9650-0805-01 Rev. A Propaq MD Operator’s Guide 13-7Locating the V1/C1 position (fourth intercostal space) is critically im

Page 66 - CHAPTER 4TRENDS

CHAPTER 13 MONITORING 12-LEAD ECG13-8 www.zoll.com 9650-0805-01 Rev. AObserving the 12-Lead Waveform TracesTo observe the 12-Lead waveform traces, pr

Page 67 - Continuous Waveform Recording

Printing 12-Lead Waveform Traces9650-0805-01 Rev. A Propaq MD Operator’s Guide 13-9Printing 12-Lead Waveform TracesOnce you have determined that the P

Page 68

CHAPTER 1GENERAL INFORMATION1-6 www.zoll.com 9650-0805-01 Rev. AConventionsThis guide uses the following conventions:Within text, the names and label

Page 69 - Chapter 5

CHAPTER 13 MONITORING 12-LEAD ECG13-10 www.zoll.com 9650-0805-01 Rev. A12-Lead Print and Display OptionsThe Propaq MD unit provides additional 12-Lea

Page 70 - Alarm Indicator Self-Test

12-Lead Print and Display Options9650-0805-01 Rev. A Propaq MD Operator’s Guide 13-11Specifying the 12-Lead Print FormatThis option allows you to spec

Page 71 - Equipment Alert Display

CHAPTER 13 MONITORING 12-LEAD ECG13-12 www.zoll.com 9650-0805-01 Rev. A

Page 72

9650-0805-01 Rev. A Propaq MD Operator’s Guide 14–1Chapter 14Manual DefibrillationEmergency Defibrillation Procedure with PaddlesWarning! To avoid ris

Page 73 - Suspending Alarms

MANUAL DEFIBRILLATION14-2 www.zoll.com 9650-0805-01 Rev. ADetermine the Patient’s Condition Following Local Medical ProtocolsVerify the following:• Un

Page 74

Emergency Defibrillation Procedure with Paddles9650-0805-01 Rev. A Propaq MD Operator’s Guide 14–3Apply Paddles to ChestApply the paddles firmly to th

Page 75 - Alarm Options

MANUAL DEFIBRILLATION14-4 www.zoll.com 9650-0805-01 Rev. AA charging message displays at the bottom of the screen, and a distinctive charging tone sou

Page 76 - CHAPTER 5ALARMS

Emergency Defibrillation Procedure with Hands-Free Therapy Electrodes9650-0805-01 Rev. A Propaq MD Operator’s Guide 14–5Note: The front panel Shock bu

Page 77

MANUAL DEFIBRILLATION14-6 www.zoll.com 9650-0805-01 Rev. ABegin CPR Following Medical ProtocolsRequest additional assistance.Prepare PatientRemove all

Page 78

Emergency Defibrillation Procedure with Hands-Free Therapy Electrodes9650-0805-01 Rev. A Propaq MD Operator’s Guide 14–71 Select Energy LevelPress the

Page 79 - Monitoring ECG

Propaq MD Indications for Use9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-7When the pediatric patient is less than 8 years of age or weighs less t

Page 80 - ECG Monitoring Setup

MANUAL DEFIBRILLATION14-8 www.zoll.com 9650-0805-01 Rev. AA charging message displays at the bottom of the screen, and a distinctive charging tone sou

Page 81

Internal Paddles9650-0805-01 Rev. A Propaq MD Operator’s Guide 14–9Note: If at any time you want to cancel the defibrillation, press the Disarm softke

Page 82 - 5-Lead Electrode Placement

MANUAL DEFIBRILLATION14-10 www.zoll.com 9650-0805-01 Rev. A7. Discharge the energy in the following manner.• For internal handles with a Discharge but

Page 83

Synchronized Cardioversion Procedure9650-0805-01 Rev. A Propaq MD Operator’s Guide 14–11Synchronized Cardioversion ProcedureDetermine the Patient’s Co

Page 84

MANUAL DEFIBRILLATION14-12 www.zoll.com 9650-0805-01 Rev. ANote: The unit can be configured to stay in SYNC mode after defibrillation inDefib/Pacer de

Page 85

Synchronized Cardioversion Procedure9650-0805-01 Rev. A Propaq MD Operator’s Guide 14–133 Deliver ShockWarning! Warn all persons in attendance of the

Page 86

MANUAL DEFIBRILLATION14-14 www.zoll.com 9650-0805-01 Rev. A

Page 87 - ECG System Messages

9650-0805-01 Rev. A Propaq MD Operator’s Guide 15–1Chapter 15External PacingWarning! Pacing is intended for use on adult patients and on adolescent, c

Page 88

EXTERNAL PACING15-2 www.zoll.com 9650-0805-01 Rev. AExternal PacingPropaq MD defibrillators contain a pacemaker for the emergency treatment of hemodyn

Page 89 - Chapter 7

Pacing in Demand Mode9650-0805-01 Rev. A Propaq MD Operator’s Guide 15–3patient). Attach hands-free therapy electrodes according to instructions on th

Page 90 - Respiration/Breath Rate Meter

CHAPTER 1GENERAL INFORMATION1-8 www.zoll.com 9650-0805-01 Rev. ATemperature MonitoringThe Propaq MD is intended for use to make continuous temperatur

Page 91

EXTERNAL PACING15-4 www.zoll.com 9650-0805-01 Rev. ANote: The initial pacing mode can be set in Defib/Pacer Default settings in the Supervisor Setup m

Page 92 - Respiration Rate Alarm Limits

Pacing in Fixed Mode9650-0805-01 Rev. A Propaq MD Operator’s Guide 15–5Warning! Determination of electrical capture should only be performed by viewin

Page 93 - Heart Rate Meter

EXTERNAL PACING15-6 www.zoll.com 9650-0805-01 Rev. ATherapy Electrode ApplicationWarning! Poor adherence and/or air under the therapy electrodes can l

Page 94

Pacing in Fixed Mode9650-0805-01 Rev. A Propaq MD Operator’s Guide 15–75 Set Pacer RateUse the arrow keys to navigate to Rate, press the Select button

Page 95

EXTERNAL PACING15-8 www.zoll.com 9650-0805-01 Rev. AMechanical capture is assessed by palpation of the peripheral pulse.To avoid mistaking muscular re

Page 96

9650-0805-01 Rev. A Propaq MD Operator’s Guide 16–1Chapter 16Patient DataThis chapter describes procedures for storing, viewing, and transferring pati

Page 97 - Chapter 8

PATIENT DATA16–2 www.zoll.com 9650-0805-01 Rev. ACapturing a Data SnapshotPress the Snapshot button ( ) on the front panel to capture a 24-second peri

Page 98 - How does NIBP Work?

Transferring Data to a USB Device9650-0805-01 Rev. A Propaq MD Operator’s Guide 16–3Printing Treatment Summary ReportTo print a treatment summary repo

Page 99 - NIBP Setup and Use

PATIENT DATA16–4 www.zoll.com 9650-0805-01 Rev. AAfter transferring data, you must remove the USB drive and reinsert it before attempting another tran

Page 100 - Selecting the NIBP Cuff

9650-0805-01 Rev. A Propaq MD Operator’s Guide 17–1Chapter 17PrintingThis chapter describes procedures for printing permanent records of pertinent pat

Page 101

Propaq MD Product Functions9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-9Propaq MD Product FunctionsDefibrillator FunctionThe Propaq MD contains a

Page 102 - Connecting the NIBP Cuff

CHAPTER 17 PRINTING17–2 www.zoll.com 9650-0805-01 Rev. APrinter SetupTo configure the printer functions, press the More quick access key ( ), press th

Page 103

Printing Patient Data9650-0805-01 Rev. A Propaq MD Operator’s Guide 17–3The SpO2 PlethysmographThe SpO2 plethysmograph is printed with major divisions

Page 104

CHAPTER 17 PRINTING17–4 www.zoll.com 9650-0805-01 Rev. APrinting TrendsPrinted trends are useful for reviewing the patient’s vital signs over the last

Page 105

9650-0805-01 Rev. A Propaq MD Operator’s Guide 18-1Chapter 18MaintenanceResuscitation equipment must be maintained to be ready for immediate use. To e

Page 106

CHAPTER 18 MAINTENANCE18-2 www.zoll.com 9650-0805-01 Rev. ADaily/Shift Check ProcedureInspectionEquipment and Accessories• Ensure that the Propaq MD

Page 107

Defibrillator/Pacing Test with Hands-Free Therapy Electrodes9650-0805-01 Rev. A Propaq MD Operator’s Guide 18-3Defibrillator/Pacing Test with Hands-Fr

Page 108 - 200 250 300

CHAPTER 18 MAINTENANCE18-4 www.zoll.com 9650-0805-01 Rev. A8 Press the Select Energy arrow (up or down). Verify that the defibrillation window displa

Page 109 - Specify NIBP Mode

Defibrillator Testing with External Paddles9650-0805-01 Rev. A Propaq MD Operator’s Guide 18-5Defibrillator Testing with External PaddlesPrior to test

Page 110 - Start/Stop TurboCuf

CHAPTER 18 MAINTENANCE18-6 www.zoll.com 9650-0805-01 Rev. A12 While the defibrillator is still charged, press the paddles electrodes together then pr

Page 111 - NIBP System Messages

Recommended Minimum Preventive Maintenance Schedule9650-0805-01 Rev. A Propaq MD Operator’s Guide 18-7Recommended Minimum Preventive Maintenance Sched

Page 112

The issue date for the Propaq MD Operator's Guide (REF 9650-0805-01 Rev. A) is March, 2012.If more than 3 years have elapsed since the issue date

Page 113 - Chapter 9

CHAPTER 1GENERAL INFORMATION1-10 www.zoll.com 9650-0805-01 Rev. AElectrodesThe Propaq MD will defibrillate, cardiovert, and monitor ECG using hands-f

Page 114 - , Oximeter Sensor

CHAPTER 18 MAINTENANCE18-8 www.zoll.com 9650-0805-01 Rev. ACleaning instructionsCleaning the Propaq MD unitTo clean the Propaq MD unit, use a nearly

Page 115 - Setup and Use

Cleaning instructions9650-0805-01 Rev. A Propaq MD Operator’s Guide 18-9Cleaning SpO2 SensorsClean the SpO2 sensors with a cloth that has been slightl

Page 116 - Applying a Reusable SpO

CHAPTER 18 MAINTENANCE18-10 www.zoll.com 9650-0805-01 Rev. A2. Refer to the illustration inside of the paper compartment for proper paper orientation

Page 117 - Applying a Single-Use Sensor

9650-0805-01 Rev. A Propaq MD Operator’s Guide 18-11Signature _______________________________________________Propaq MD OPERATOR’S SHIFT CHECKLISTDate:

Page 118 - Connecting the SpO

CHAPTER 18 MAINTENANCE18-12 www.zoll.com 9650-0805-01 Rev. A

Page 119 - Enabling/Disabling SpO

9650-0805-01 Rev. A Propaq MD Operator’s Guide A-1Appendix ASpecificationsThis chapter provides specification information for the Propaq MD Monitor/De

Page 120 - Parameter Control Panel

SpecificationsA-2 www.zoll.com 9650-0805-01 Rev. ADefibrillatorCharge Time:• Less than 7 seconds with a new, fully charged battery (first 15 charges

Page 121 - System Messages

Defibrillator9650-0805-01 Rev. A Propaq MD Operator’s Guide A-3* For all energy levels, accuracy is equal to either ±15% or 3 joules, whichever is gre

Page 122

SpecificationsA-4 www.zoll.com 9650-0805-01 Rev. AThe vertical axis shows the current in amperes (A); the horizontal axis shows the duration of time

Page 123 - Monitoring CO

Defibrillator9650-0805-01 Rev. A Propaq MD Operator’s Guide A-5Figure A-3. Rectilinear Biphasic Waveform at 120 JoulesFigure A-4. Rectilinear Biphasic

Page 124 - Monitoring Setup and Use

Propaq MD Product Functions9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-11• External charging — use the ZOLL SurePower™ Battery Charger with the P

Page 125 - Sampling Line

SpecificationsA-6 www.zoll.com 9650-0805-01 Rev. AFigure A-5. Rectilinear Biphasic Waveform at 85 JoulesFigure A-6. Rectilinear Biphasic Waveform at

Page 126 - Sampling Lines

Defibrillator9650-0805-01 Rev. A Propaq MD Operator’s Guide A-7Figure A-7. Rectilinear Biphasic Waveform at 50 JoulesFigure A-8. Rectilinear Biphasic

Page 127 - Applying a FilterLine Set

SpecificationsA-8 www.zoll.com 9650-0805-01 Rev. AFigure A-9. Rectilinear Biphasic Waveform at 20 JoulesFigure A-10. Rectilinear Biphasic Waveform at

Page 128 - CHAPTER 10 MONITORING CO2

Defibrillator9650-0805-01 Rev. A Propaq MD Operator’s Guide A-9Figure A-11. Rectilinear Biphasic Waveform at 10 JoulesFigure A-12. Rectilinear Biphasi

Page 129 - Measuring CO

SpecificationsA-10 www.zoll.com 9650-0805-01 Rev. AFigure A-13. Rectilinear Biphasic Waveform at 8 JoulesFigure A-14. Rectilinear Biphasic Waveform a

Page 130 - and Respiration Rate Alarms

Defibrillator9650-0805-01 Rev. A Propaq MD Operator’s Guide A-11Figure A-15. Rectilinear Biphasic Waveform at 6 JoulesFigure A-16. Rectilinear Biphasi

Page 131 - Setting Upper FiCO

SpecificationsA-12 www.zoll.com 9650-0805-01 Rev. AFigure A-17. Rectilinear Biphasic Waveform at 4 JoulesFigure A-18. Rectilinear Biphasic Waveform a

Page 132

Defibrillator9650-0805-01 Rev. A Propaq MD Operator’s Guide A-13Figure A-19. Rectilinear Biphasic Waveform at 2 JoulesFigure A-20. Rectilinear Biphasi

Page 133 - System Messages

SpecificationsA-14 www.zoll.com 9650-0805-01 Rev. AMonitor/DisplayInput: 3-lead, 5-lead, or 12-lead patient cable, paddles, or Multifunction electrod

Page 134 - NO IMPLIED LICENSE

Impedance Pneumography9650-0805-01 Rev. A Propaq MD Operator’s Guide A-15Heart Rate Response to Irregular Rhythm: (AAMI EC13-2002, section 4.1.2.1.e.)

Page 135 - Chapter 11

CHAPTER 1GENERAL INFORMATION1-12 www.zoll.com 9650-0805-01 Rev. AReady For Use (RFU) IndicatorThe Propaq MD has an RFU indicator on the front panel t

Page 136 - IBP Setup

SpecificationsA-16 www.zoll.com 9650-0805-01 Rev. AAlarmsHeart Rate Alarms:Audible: 5 pulse, 900 Hz tone, with a PW of 125 msec, a PRI of 250 msec, a

Page 137 - Zeroing the Transducer

Recorder9650-0805-01 Rev. A Propaq MD Operator’s Guide A-17RecorderType: High-resolution thermal array.Annotation: Time, date, ECG lead, ECG gain, hea

Page 138 - Rezeroing a Transducer

SpecificationsA-18 www.zoll.com 9650-0805-01 Rev. AGeneralWeight: 9.9 lbs.without battery and paper11.6 lbs.with battery and paperDimensions: Without

Page 139 - Displaying IBP Measurements

Pacer9650-0805-01 Rev. A Propaq MD Operator’s Guide A-19PacerType: External transcutaneous pacingPacer Rate: 30 to 180 BPM ± 1.5%.Output Current: 0 to

Page 140

SpecificationsA-20 www.zoll.com 9650-0805-01 Rev. ARespiration Rate Accuracy: 0 to 70 bpm: ±1 bpm71 to 120 bpm: ±2 bpm121 to 149 bpm: ±3 bpmFlow

Page 141

Non-Invasive Blood Pressure9650-0805-01 Rev. A Propaq MD Operator’s Guide A-21Non-Invasive Blood PressureTechnique: Non-invasive oscillometric methodO

Page 142 - Setting IBP Source Label

SpecificationsA-22 www.zoll.com 9650-0805-01 Rev. ABlood Pressure Validation:Blood pressure measurements determined with this device are equivalent t

Page 143 - IBP System Messages

Clinical Trial Results for the Biphasic Waveform9650-0805-01 Rev. A Propaq MD Operator’s Guide A-23Clinical Trial Results for the Biphasic WaveformThe

Page 144

SpecificationsA-24 www.zoll.com 9650-0805-01 Rev. ASuccessful defibrillation with rectilinear biphasic shocks was achieved with 58% less delivered cu

Page 145 - Monitoring Temperature

Clinical Trial Results for the Biphasic Waveform9650-0805-01 Rev. A Propaq MD Operator’s Guide A-25The total efficacy of consecutive rectilinear bipha

Page 146 - Displaying Temperature

Safety Considerations9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-13Safety ConsiderationsAll operators should review these safety considerations b

Page 147 - 2. Press

SpecificationsA-26 www.zoll.com 9650-0805-01 Rev. ASynchronized Cardioversion of Atrial FibrillationCardioversion of atrial fibrillation (AF) and ove

Page 148

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration9650-0805-01 Rev. A Propaq MD Operator’s Guide A-27Electromagnetic Compatibility

Page 149 - Temperature System Messages

SpecificationsA-28 www.zoll.com 9650-0805-01 Rev. AGuidance and Manufacturer’s Declaration – Electromagnetic(IEC 60601-1-2 Table 202)The Propaq MD un

Page 150

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration9650-0805-01 Rev. A Propaq MD Operator’s Guide A-29Guidance and Manufacturer’s De

Page 151 - Monitoring 12-Lead ECG

SpecificationsA-30 www.zoll.com 9650-0805-01 Rev. ARecommended Separation Distances Between Portable and Mobile RF Communications Equipment and the P

Page 152 - Entering Patient Information

9650-0805-01 Rev. A Propaq MD Operator’s Guide B-1Appendix BAccessoriesThe following accessories are compatible for use with the Propaq MD. To order a

Page 153

ACCESSORIESB-2 www.zoll.com 9650-0805-01 Rev. AFilterLine H Set, Adult/Pediatric, box of 25FilterLine H Set, Infant/Neonate, box of 25FilterLine Set,

Page 154 - 12-Lead ECG Monitoring Setup

9650-0805-01 Rev. A Propaq MD Operator’s Guide B-3YSI Reusable Adult, Esophageal/RectalYSI Reusable Pediatric, Esophageal/RectalSensor Adapter Cable f

Page 155

ACCESSORIESB-4 www.zoll.com 9650-0805-01 Rev. AOther AccessoriesPropaq MD Molded Carrying CaseECG 50mm Chart Recorder Paper

Page 156

CHAPTER 1GENERAL INFORMATION1-14 www.zoll.com 9650-0805-01 Rev. AThe Propaq MD unit might not perform to specifications when stored at the upper or l

Page 157 - Connecting the 12-Lead Cable

Warnings9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-15Equipment such as electrocautery or diathermy equipment, RFID readers, electronic article s

Page 158 - 00:17:43

CHAPTER 1GENERAL INFORMATION1-16 www.zoll.com 9650-0805-01 Rev. ABefore charging the defibrillator, verify that the energy selected on the display is

Page 159

Warnings9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-17Transcutaneous pacing may cause discomfort ranging from mild to severe, depending on the pa

Page 160 - Selecting 12-Lead Acquire

CHAPTER 1GENERAL INFORMATION1-18 www.zoll.com 9650-0805-01 Rev. ANoninvasive Blood PressureOnly a physician can interpret pressure measurements.Blood

Page 161 - Diagnostic 0.05 to 150 Hz

Warnings9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-19CO2During MRI scanning, the monitor must be placed outside the MRI suite. When the monitor

Page 162

9650-0805-01 Rev. A Propaq MD Operator’s Guide iTable of ContentsChapter 1 General InformationProduct Description ...

Page 163 - Manual Defibrillation

CHAPTER 1GENERAL INFORMATION1-20 www.zoll.com 9650-0805-01 Rev. AIf the LOW BATTERY icon appears, plug the Propaq MD unit into a power source or inst

Page 164 - 1 Select Energy Level

Warnings9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-21Do not touch the bed, patient, or any equipment connected to the patient during defibrillat

Page 165 - 2 Charge Defibrillator

CHAPTER 1GENERAL INFORMATION1-22 www.zoll.com 9650-0805-01 Rev. ADuring electrosurgery, observe the following guidelines to minimize electrosurgery u

Page 166 - Press to Shock

Restarting the Defibrillator9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-23Restarting the DefibrillatorCertain events require the Propaq MD produc

Page 167 - Therapy Electrodes

CHAPTER 1GENERAL INFORMATION1-24 www.zoll.com 9650-0805-01 Rev. ANotification of Adverse EventsAs a health care provider, you may have responsibiliti

Page 168 - Prepare Patient

Service9650-0805-01 Rev. A Propaq MD Operator’s Guide 1-25ServiceThe Propaq MD only requires recalibration of the CO2 module. Service is required afte

Page 169

CHAPTER 1GENERAL INFORMATION1-26 www.zoll.com 9650-0805-01 Rev. AReturning a unit for serviceBefore sending a unit to the ZOLL Technical Service Depa

Page 170 - 3 Deliver Shock

9650-0805-01 Rev. A Propaq MD Operator’s Guide 2–1Chapter 2Product OverviewDefibrillator Controls and Indicators52916457835

Page 171 - Internal Paddles

CHAPTER 2PRODUCT OVERVIEW2–2 www.zoll.com 9650-0805-01 Rev. AThe Front PanelThe front panel of the Propaq MD device includes the display screen, quick

Page 172 - Synchronized Cardioversion

Defibrillator Controls and Indicators9650-0805-01 Rev. A Propaq MD Operator’s Guide 2–3Table 2-2. Propaq Controls and IndicatorsControl or Indicator D

Page 173 - Press the Sync Key

TABLE OF CONTENTSii www.zoll.com 9650-0805-01 Rev. ARestarting the Defibrillator ...

Page 174

CHAPTER 2PRODUCT OVERVIEW2–4 www.zoll.com 9650-0805-01 Rev. ADisplay ScreenThe front panel includes a color display which shows:• Date and time• Patie

Page 175

Defibrillator Controls and Indicators9650-0805-01 Rev. A Propaq MD Operator’s Guide 2–5• EtCO2 numeric data• Invasive pressure numeric data• The selec

Page 176 - MANUAL DEFIBRILLATION

CHAPTER 2PRODUCT OVERVIEW2–6 www.zoll.com 9650-0805-01 Rev. ABattery Status and Auxiliary Power IndicatorsThe battery status indicator displays variou

Page 177 - External Pacing

Defibrillator Controls and Indicators9650-0805-01 Rev. A Propaq MD Operator’s Guide 2–7Patient Cables and ConnectorsThe left and right sides of the un

Page 178 - Pacing in Demand Mode

CHAPTER 2PRODUCT OVERVIEW2–8 www.zoll.com 9650-0805-01 Rev. AMultifunction Cable (MFC)The unit ships with an MFC that is used to defibrillate the pati

Page 179 - 4 Set Mode

Defibrillator Controls and Indicators9650-0805-01 Rev. A Propaq MD Operator’s Guide 2–9Removing MFC from UnitTwist the connector to the left to unlock

Page 180 - 8 Determine Capture

CHAPTER 2PRODUCT OVERVIEW2–10 www.zoll.com 9650-0805-01 Rev. AFigure 2-7. MFC Connected to APEX PaddleIf you need to detach the MFC from the APEX pad

Page 181 - Pacing in Fixed Mode

Defibrillator Controls and Indicators9650-0805-01 Rev. A Propaq MD Operator’s Guide 2–11Pediatric-size electrodes are built in to the paddle assembly

Page 182

CHAPTER 2PRODUCT OVERVIEW2–12 www.zoll.com 9650-0805-01 Rev. ATo connect the auxiliary power adapter, align the white arrow of the auxiliary power con

Page 183

Navigating the Display Screen9650-0805-01 Rev. A Propaq MD Operator’s Guide 2–13Table 2-3. Propaq MD Quick Access KeysQuick access key DescriptionLead

Page 184 - Pediatric Pacing

9650-0805-01 Rev. A Propaq MD Operator’s Guide iiiChapter 5 AlarmsVisual Alarm Indicators ...

Page 185 - Patient Data

CHAPTER 2PRODUCT OVERVIEW2–14 www.zoll.com 9650-0805-01 Rev. ATransfer Log Transfers the current data in the log to a USB drive.Clear Log Deletes the

Page 186 - Capturing a Data Snapshot

Navigating the Display Screen9650-0805-01 Rev. A Propaq MD Operator’s Guide 2–15Navigation Keys Use the navigation keys (up/clockwise arrow, down/coun

Page 187 - USB Device Port

CHAPTER 2PRODUCT OVERVIEW2–16 www.zoll.com 9650-0805-01 Rev. ACommon TasksThe section contains procedures for the following tasks:• “Changing the Disp

Page 188 - Clearing the Log

Common Tasks9650-0805-01 Rev. A Propaq MD Operator’s Guide 2–172. Push the battery into place.Figure 2-13. Installing a Battery PackUsing Treatment Bu

Page 189 - Printing

CHAPTER 2PRODUCT OVERVIEW2–18 www.zoll.com 9650-0805-01 Rev. A

Page 190 - Printing Waveforms

9650-0805-01 Rev. A Propaq MD Operator’s Guide 3-1Chapter 3Monitoring OverviewThis chapter provides an overview of the Propaq MD unit’s monitoring fun

Page 191 - Printing Reports

CHAPTER 3MONITORING OVERVIEW3-2 www.zoll.com 9650-0805-01 Rev. AECG An ECG waveform trace appears at the top of the display area. You can specify tha

Page 192 - Printing Trends

Monitoring Display Options9650-0805-01 Rev. A Propaq MD Operator’s Guide 3-3Capnography (CO2) CO2 monitoring measures the CO2 concentration in a patie

Page 193 - Maintenance

CHAPTER 3MONITORING OVERVIEW3-4 www.zoll.com 9650-0805-01 Rev. AYou can display up to four waveform traces that you specify on the Waveform Display w

Page 194 - Daily/Shift Check Procedure

Monitoring Display Options9650-0805-01 Rev. A Propaq MD Operator’s Guide 3-5 03/23/2009 12:34:56Adult00:17:43I, II,III...IBP12CO2Analyze1 cm/mV

Page 195 - Electrodes

TABLE OF CONTENTSiv www.zoll.com 9650-0805-01 Rev. AApplying the Cuff to the Patient ...

Page 196 - CHAPTER 18 MAINTENANCE

CHAPTER 3MONITORING OVERVIEW3-6 www.zoll.com 9650-0805-01 Rev. APress the Home/Display button when viewing the Trends Status window and the Large Num

Page 197

Configuring the Waveform Display9650-0805-01 Rev. A Propaq MD Operator’s Guide 3-7Configuring the Waveform DisplayYou can display up to four waveform

Page 198

CHAPTER 3MONITORING OVERVIEW3-8 www.zoll.com 9650-0805-01 Rev. AWhen the unit cascades the ECG Lead lI trace, the Waveform Display window appears as

Page 199 - Annually

Configuring the Waveform Display9650-0805-01 Rev. A Propaq MD Operator’s Guide 3-9Inserting a third waveform trace for ECG lead aVR:03/23/2009

Page 200 - Cleaning instructions

CHAPTER 3MONITORING OVERVIEW3-10 www.zoll.com 9650-0805-01 Rev. AInserting a capnogram (CO2) into the fourth trace area: 03/23/2009 12:34:56 A

Page 201 - Loading Recorder Paper

9650-0805-01 Rev. A Propaq MD Operator’s Guide 4-1Chapter 4TrendsThe Propaq MD unit accumulates a patient’s trend information by logging all monitored

Page 202 - Cleaning the Print Head

CHAPTER 4TRENDS4-2 www.zoll.com 9650-0805-01 Rev. AFigure 4-1 Trends Status windowThe Trends status window displays the logged trend information and

Page 203

Changing the Trends Status Window Display9650-0805-01 Rev. A Propaq MD Operator’s Guide 4-3Changing the Trends Status Window DisplayBy default, the Tr

Page 204

CHAPTER 4TRENDS4-4 www.zoll.com 9650-0805-01 Rev. A

Page 205 - Specifications

9650-0805-01 Rev. A Propaq MD Operator’s Guide 5-1Chapter 5AlarmsThe Propaq MD unit supports the detection and indication of patient alarms and techni

Page 206 - Defibrillator

9650-0805-01 Rev. A Propaq MD Operator’s Guide vChapter 11 Monitoring Invasive Pressures (IBP)Invasive Pressure Transducers ...

Page 207

CHAPTER 5ALARMS5-2 www.zoll.com 9650-0805-01 Rev. AAudible Alarm IndicatorsThe Propaq MD unit sounds an audible alarm to indicate the priority level

Page 208

Patient Alarm Display9650-0805-01 Rev. A Propaq MD Operator’s Guide 5-3Patient Alarm DisplayWhen a patient’s vital signs measurements trigger an alarm

Page 209

CHAPTER 5ALARMS5-4 www.zoll.com 9650-0805-01 Rev. AIn the following example, an equipment alert message indicates that the SpO2 sensor has become una

Page 210

Suspending Alarms9650-0805-01 Rev. A Propaq MD Operator’s Guide 5-5Re-enabling an AlarmTo re-enable an alarm before the alarm silence period has expir

Page 211

CHAPTER 5ALARMS5-6 www.zoll.com 9650-0805-01 Rev. AThe Alarm Suspension TimerDuring an alarm suspension, the window displays an alarm suspension time

Page 212

Alarm Options9650-0805-01 Rev. A Propaq MD Operator’s Guide 5-7Alarm OptionsThe Propaq MD unit provides alarm options that you can specify through the

Page 213

CHAPTER 5ALARMS5-8 www.zoll.com 9650-0805-01 Rev. ASelecting Default Alarm LimitsThe first three options -- Default Adult, Default Pediatric, Default

Page 214

Alarm Options9650-0805-01 Rev. A Propaq MD Operator’s Guide 5-9SpO2(%)Entire range Limit = 100 (Adult and Pediatric)Limit = Numeric + 5 (Neo-nate)Limi

Page 215

CHAPTER 5ALARMS5-10 www.zoll.com 9650-0805-01 Rev. A

Page 216

9650-0805-01 Rev. A Propaq MD Operator’s Guide 6-1Chapter 6Monitoring ECGThis chapter describes how to use the Propaq MD unit to monitor ECG.Propaq MD

Page 217

TABLE OF CONTENTSvi www.zoll.com 9650-0805-01 Rev. AChapter 14 Manual DefibrillationEmergency Defibrillation Procedure with Paddles...

Page 218 - Monitor/Display

CHAPTER 6MONITORING ECG6-2 www.zoll.com 9650-0805-01 Rev. A• To avoid a shock hazard and interference from nearby electrical equipment, keep electrod

Page 219 - Impedance Pneumography

ECG Monitoring Setup9650-0805-01 Rev. A Propaq MD Operator’s Guide 6-3Applying Electrodes to the PatientThe following sections show where to place ele

Page 220 - • if source is IBP, 5 seconds

CHAPTER 6MONITORING ECG6-4 www.zoll.com 9650-0805-01 Rev. A5-Lead Electrode PlacementDepending upon local usage, the ECG leads are marked either RA,

Page 221 - Battery

ECG Monitoring Setup9650-0805-01 Rev. A Propaq MD Operator’s Guide 6-5Figure 6-2 5-Lead Electrode PlacementConnecting the ECG Cable To the Propaq MD U

Page 222 - • EN 1789 for ambulance

CHAPTER 6MONITORING ECG6-6 www.zoll.com 9650-0805-01 Rev. ASelecting ECG Waveforms for DisplayYou can fit up to four waveforms on the Propaq MD displ

Page 223 - Accuracy CO

ECG Monitoring Setup9650-0805-01 Rev. A Propaq MD Operator’s Guide 6-7Select a current waveform source, and the unit displays the waveform immediately

Page 224 - Pulse Oximeter

CHAPTER 6MONITORING ECG6-8 www.zoll.com 9650-0805-01 Rev. ASelecting the Waveform Trace SizeThe Propaq MD unit allows you to select the waveform trac

Page 225 - Non-Invasive Blood Pressure

ECG Monitoring and Pacemakers9650-0805-01 Rev. A Propaq MD Operator’s Guide 6-9ECG Monitoring and PacemakersWhen the unit performs ECG monitoring on a

Page 226 - Temperature

CHAPTER 6MONITORING ECG6-10 www.zoll.com 9650-0805-01 Rev. A

Page 227 - First shock efficacy 93% 99%

9650-0805-01 Rev. A Propaq MD Operator’s Guide 7-1Chapter 7Monitoring Respiration (Resp) and Heart Rate (HR)This chapter describes how to use the Prop

Page 228 - Atrial Fibrillation (AF)

9650-0805-01 Rev. A Propaq MD Operator’s Guide viiPacing in Fixed Mode ...

Page 229

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)7-2 www.zoll.com 9650-0805-01 Rev. ARespiration/Breath Rate MeterIf enabled, the Propaq MD

Page 230 - Posterior

Configuring Respiration (RR/BR) Alarms and Settings9650-0805-01 Rev. A Propaq MD Operator’s Guide 7-3Configuring Respiration (RR/BR) Alarms and Settin

Page 231 - Manufacturer’s Declaration

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)7-4 www.zoll.com 9650-0805-01 Rev. ARespiration Rate Alarm LimitsInitially, the Resp Alarm

Page 232 - (IEC 60601-1-2 Table 202)

Heart Rate Meter9650-0805-01 Rev. A Propaq MD Operator’s Guide 7-5Enabling/Disabling Resp MonitoringSelect the Resp Monitoring prompt to enable or dis

Page 233 - 60601-1-2 Table 203)

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)7-6 www.zoll.com 9650-0805-01 Rev. AEnabling/Disabling HR Alarms and Setting Alarm LimitsW

Page 234

Configuring Heart Rate (HR) Meter Alarms9650-0805-01 Rev. A Propaq MD Operator’s Guide 7-7Using the Heart Rate Parameter Control PanelTo configure ala

Page 235 - Accessories

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)7-8 www.zoll.com 9650-0805-01 Rev. A

Page 236 - ACCESSORIES

9650-0805-01- Rev. A Propaq MD Operator’s Guide 8-1Chapter 8Monitoring Non-Invasive Blood Pressure (NIBP)The Propaq MD NIBP input is Type CF defibrill

Page 237

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-2 www.zoll.com 9650-0805-01- Rev. A• Do not use NIBP without proper training. Check that the

Page 238

The NIBP Numeric Display9650-0805-01- Rev. A Propaq MD Operator’s Guide 8-33. The NIBP option automatically adjusts the blood pressure measurement pro