Zoll Propaq MD Rev C User Manual

Browse online or download User Manual for Health and hygiene Zoll Propaq MD Rev C. ZOLL Propaq MD Rev C User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 326
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - MD Operator’s Guide

Propaq® MD Operator’s Guide9650-0806-01 Rev. C

Page 2

TABLE OF CONTENTSviii www.zoll.com 9650-0806-01 Rev. CChapter 18 External PacingExternal Pacing...

Page 3 - Table of Contents

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)7-6 www.zoll.com 9650-0806-01 Rev. CEnabling/Disabling HR Alarms and Setting Alarm LimitsW

Page 4 - Chapter 2 Product Overview

Configuring Heart Rate (HR) Meter Alarms9650-0806-01 Rev. C Propaq MD Operator’s Guide 7-7Life Threatening Rhythm AlarmsWhen LTA monitoring is enabled

Page 5 - Chapter 6 Monitoring ECG

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)7-8 www.zoll.com 9650-0806-01 Rev. CExtreme Bradycardia and Extreme Tachycardia Alarm Limi

Page 6 - Chapter 9 Monitoring CO

RESP System Message9650-0806-01 Rev. C Propaq MD Operator’s Guide 7-9Using the Heart Rate Parameter Control PanelTo configure alarms through the Heart

Page 7

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)7-10 www.zoll.com 9650-0806-01 Rev. C

Page 8 - ABLE OF CONTENTS

9650-0806-01 Rev. C Propaq MD Operator’s Guide 8-1Chapter 8Monitoring Non-Invasive Blood Pressure (NIBP)The Propaq MD NIBP input is Type CF defibrilla

Page 9

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-2 www.zoll.com 9650-0806-01 Rev. C• Do not use NIBP without proper training. Check that the c

Page 10 - Chapter 20 See-Thru CPR

The NIBP Numeric Display9650-0806-01 Rev. C Propaq MD Operator’s Guide 8-33. The NIBP option automatically adjusts the blood pressure measurement proc

Page 11 - Chapter 24 Maintenance

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-4 www.zoll.com 9650-0806-01 Rev. CSelecting the NIBP CuffTo take accurate measurements, you m

Page 12 - Appendix B Accessories

Connecting the NIBP Cuff9650-0806-01 Rev. C Propaq MD Operator’s Guide 8-5Connecting the NIBP CuffConnecting the NIBP cuff requires you to attach the

Page 13 - General Information

9650-0806-01 Rev. C Propaq MD Operator’s Guide ixChapter 21 Patient DataStoring Data ...

Page 14 - Propaq MD Optional Features

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-6 www.zoll.com 9650-0806-01 Rev. CAttach a single lumen hose as follows:Figure 8-2 Attaching

Page 15 - How to Use This Manual

Applying the Cuff to the Patient9650-0806-01 Rev. C Propaq MD Operator’s Guide 8-72. Insert the plastic connectors on the NIBP hose into the cuff hose

Page 16 - Symbols Used on the Equipment

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-8 www.zoll.com 9650-0806-01 Rev. CCaution • Using a cuff that is loosely applied or too small

Page 17

Configuring NIBP Alarms and Settings9650-0806-01 Rev. C Propaq MD Operator’s Guide 8-9In order to accurately measure systolic pressure, the CIT pressu

Page 18 - CHAPTER 1GENERAL INFORMATION

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-10 www.zoll.com 9650-0806-01 Rev. C5. On the selected NIBP alarm settings menu, use the navig

Page 19 - Conventions

Configuring NIBP Alarms and Settings9650-0806-01 Rev. C Propaq MD Operator’s Guide 8-11Setting Upper and Lower NIBP MAP Alarm LimitsInitially, the NIB

Page 20 - Semiautomatic Operation (AED)

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-12 www.zoll.com 9650-0806-01 Rev. COn the NIBP Parameter Control Panel, you can select an ala

Page 21 - Propaq MD Indications for Use

Configuring NIBP Alarms and Settings9650-0806-01 Rev. C Propaq MD Operator’s Guide 8-13By default, Smartcuf is Enabled.There are situations where you

Page 22 - Propaq MD Product Functions

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-14 www.zoll.com 9650-0806-01 Rev. CNIBP System MessagesWhen monitoring NIBP, the Propaq MD un

Page 23

NIBP System Messages9650-0806-01 Rev. C Propaq MD Operator’s Guide 8-15TEMP OUT OF RANGE The NIBP module is outside of its operating temperature range

Page 24 - Batteries

TABLE OF CONTENTSx www.zoll.com 9650-0806-01 Rev. CCleaning instructions ...

Page 25

CHAPTER 8MONITORING NON-INVASIVE BLOOD PRESSURE (NIBP)8-16 www.zoll.com 9650-0806-01 Rev. C

Page 26 - Warnings

9650-0806-01 Rev. C Propaq MD Operator’s Guide 9-1Chapter 9Monitoring CO2This chapter describes how to use the Propaq MD unit to monitor End Tidal Car

Page 27

CHAPTER 9MONITORING CO29-2 www.zoll.com 9650-0806-01 Rev. CWarning! • When using a sampling line for intubated patients with a closed suction system,

Page 28 - Defibrillation

CO2 Monitoring Setup and Use9650-0806-01 Rev. C Propaq MD Operator’s Guide 9-3Selecting the CO2 Sampling LineTo select the correct CO2 sampling line,

Page 29

CHAPTER 9MONITORING CO29-4 www.zoll.com 9650-0806-01 Rev. CConnecting the CO2 Sampling LinesTo connect the FilterLine or Smart CapnoLine:1. Slide ope

Page 30 - Pulse Oximeter

CO2 Monitoring Setup and Use9650-0806-01 Rev. C Propaq MD Operator’s Guide 9-5WAR NIN G! • The exhaust port for the Microstream sidestream CO2 apparat

Page 31

CHAPTER 9MONITORING CO29-6 www.zoll.com 9650-0806-01 Rev. CApplying a Smart CapnoLine Nasal or Nasal/Oral CannulaThe nasal and nasal/oral cannulas ar

Page 32 - Ferromagnetic Equipment

Measuring CO29650-0806-01 Rev. C Propaq MD Operator’s Guide 9-7Measuring CO2Once setup is complete, press to begin monitoring CO2. The numeric CO2 d

Page 33

CHAPTER 9MONITORING CO29-8 www.zoll.com 9650-0806-01 Rev. CSetting CO2 and Respiration Rate AlarmsThe Propaq MD unit sounds alarms whenever measureme

Page 34 - Patient Safety

Setting CO2 and Respiration Rate Alarms9650-0806-01 Rev. C Propaq MD Operator’s Guide 9-9Setting Upper and Lower EtCO2 LimitsInitially, the EtCO2 Alar

Page 35 - Cautions

9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-1Chapter 1General InformationProduct DescriptionThe ZOLL® Propaq® MD unit is an easy-to-use portable

Page 36 - FDA Tracking Requirements

CHAPTER 9MONITORING CO29-10 www.zoll.com 9650-0806-01 Rev. CSetting Upper and Lower Respiratory Rate (RR/BR) LimitsInitially, the RR/BR Alarm menu sp

Page 37 - Software License

System Messages9650-0806-01 Rev. C Propaq MD Operator’s Guide 9-11Setting the CO2 Sweep SpeedThe EtCO2 sweep speed determines the X-axis scale of the

Page 38 - Returning a unit for service

CHAPTER 9MONITORING CO29-12 www.zoll.com 9650-0806-01 Rev. CPatentsThe capnography component of the Propaq MD unit is covered by one or more of the f

Page 39 - The ZOLL Serial Number

9650-0806-01 Rev. C Propaq MD Operator’s Guide 10-1Chapter 10Pulse CO-Oximetry (SpO2, SPCO, and SpMet)The Propaq MD SpO2 input is Type CF defibrillato

Page 40

CHAPTER 10 PULSE CO-OXIMETRY (SPO2, SPCO, AND SPMET)10-2 www.zoll.com 9650-0806-01 Rev. CWarnings -- SpO2, General • SpO2 style pulse oximeter measur

Page 41 - Product Overview

Warnings -- SpO2, Oximeter Sensor9650-0806-01 Rev. C Propaq MD Operator’s Guide 10-3Warnings -- SpO2, Oximeter Sensor• Use only ZOLL-approved Masimo o

Page 42 - The Front Panel

CHAPTER 10 PULSE CO-OXIMETRY (SPO2, SPCO, AND SPMET)10-4 www.zoll.com 9650-0806-01 Rev. CSpO2 Setup and UseTo take accurate SpO2 measurements using t

Page 43

Applying the SpO2 Sensor9650-0806-01 Rev. C Propaq MD Operator’s Guide 10-5Applying a Two-Piece Single-Use Sensor/CableWhen applying a single-use SpO2

Page 44 - Display Screen

CHAPTER 10 PULSE CO-OXIMETRY (SPO2, SPCO, AND SPMET)10-6 www.zoll.com 9650-0806-01 Rev. CFigure 10-1 Plug the Sensor Connectors Together5. Lower the

Page 45 - Color coding

Applying the SpO2 Sensor9650-0806-01 Rev. C Propaq MD Operator’s Guide 10-7Applying a Reusable SpO2 Sensor/CableThe following instructions describe ho

Page 46 - CHAPTER 2PRODUCT OVERVIEW

CHAPTER 1GENERAL INFORMATION1-2 www.zoll.com 9650-0806-01 Rev. CThe Propaq MD unit assists caregivers during cardiopulmonary resuscitation (CPR) by e

Page 47 - Patient Cables and Connectors

CHAPTER 10 PULSE CO-OXIMETRY (SPO2, SPCO, AND SPMET)10-8 www.zoll.com 9650-0806-01 Rev. CCleaning and Reuse of SensorsReusable sensors can be cleaned

Page 48 - 1. Insert MFC into unit

Enabling/Disabling SpO2 Alarms and Setting Alarm Limits9650-0806-01 Rev. C Propaq MD Operator’s Guide 10-9Enabling/Disabling SpO2 Alarms and Setting A

Page 49 - to the left

CHAPTER 10 PULSE CO-OXIMETRY (SPO2, SPCO, AND SPMET)10-10 www.zoll.com 9650-0806-01 Rev. CSetting Upper and Lower SpCO and SpMet Alarm LimitsThe foll

Page 50 - External Paddles

SpO2 System Messages9650-0806-01 Rev. C Propaq MD Operator’s Guide 10-11Specifying the SpO2 Averaging TimeThe Masimo SpO2 module in the Propaq MD unit

Page 51 - Attaching the OneStep Cable

CHAPTER 10 PULSE CO-OXIMETRY (SPO2, SPCO, AND SPMET)10-12 www.zoll.com 9650-0806-01 Rev. CFunctional Testers and Patient SimulatorsSome models of com

Page 52

9650-0806-01 Rev. C Propaq MD Operator’s Guide 11-1Chapter 11Monitoring Invasive Pressures (IBP)The Propaq MD IBP inputs are Type CF defibrillator pro

Page 53 - Auxiliary Power Adapter

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-2 www.zoll.com 9650-0806-01 Rev. CWarning! • If electrocautery is used, always avoid using any trans

Page 54 - Navigating the Display Screen

Zeroing the Transducer9650-0806-01 Rev. C Propaq MD Operator’s Guide 11-3Figure 11-1 Plugging the Transducer into the Propaq MD UnitWhen you plug the

Page 55

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-4 www.zoll.com 9650-0806-01 Rev. CFigure 11-2 IBP Channel Parameter Control Panel6. Select Zero Prob

Page 56

Displaying IBP Measurements9650-0806-01 Rev. C Propaq MD Operator’s Guide 11-5Displaying IBP MeasurementsAfter attaching and zeroing a transducer, the

Page 57 - Common Tasks

How to Use This Manual9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-3How to Use This ManualThe Propaq MD Operator's Guide provides information

Page 58

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-6 www.zoll.com 9650-0806-01 Rev. CEnabling/Disabling IBP Alarms and Setting Alarm LimitsWhen enabled

Page 59

Enabling/Disabling IBP Alarms and Setting Alarm Limits9650-0806-01 Rev. C Propaq MD Operator’s Guide 11-7Setting Upper and Lower Diastolic (DIA) Alarm

Page 60

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-8 www.zoll.com 9650-0806-01 Rev. CSetting IBP Source LabelIn the IBP channel menu, you can select a

Page 61 - Monitoring Overview

IBP System Messages9650-0806-01 Rev. C Propaq MD Operator’s Guide 11-9IBP System MessagesThe Propaq MD unit may display the following messages when mo

Page 62 - CHAPTER 3MONITORING OVERVIEW

CHAPTER 11 MONITORING INVASIVE PRESSURES (IBP)11-10 www.zoll.com 9650-0806-01 Rev. C

Page 63 - Monitoring Display Options

9650-0806-01 Rev. C Propaq MD Operator’s Guide 12-1Chapter 12Monitoring TemperatureThe Propaq MD Temperature inputs are Type CF defibrillator proof.Th

Page 64

CHAPTER 12 MONITORING TEMPERATURE12-2 www.zoll.com 9650-0806-01 Rev. CWarning! • The application and use of metal-jacketed temperature probes that co

Page 65 - EtCO2 mmHg

Enabling/Disabling Temperature Alarms and Setting Alarm Limits9650-0806-01 Rev. C Propaq MD Operator’s Guide 12-3Enabling/Disabling Temperature Alarms

Page 66

CHAPTER 12 MONITORING TEMPERATURE12-4 www.zoll.com 9650-0806-01 Rev. CSetting Upper and Lower Temperature Alarm LimitsInitially, the Temperature

Page 67

Temperature System Messages9650-0806-01 Rev. C Propaq MD Operator’s Guide 12-5If you don’t select a label, the Temperature channels appear with the de

Page 68

CHAPTER 1GENERAL INFORMATION1-4 www.zoll.com 9650-0806-01 Rev. CSymbols Used on the EquipmentAny or all of the following symbols may be used in this

Page 69

CHAPTER 12 MONITORING TEMPERATURE12-6 www.zoll.com 9650-0806-01 Rev. C

Page 70

9650-0806-01 Rev. C Propaq MD Operator’s Guide 13–1Chapter 13Automated External Defibrillator(AED) OperationWARNING! Use only pediatric electrodes to

Page 71 - Chapter 4

CHAPTER 13 AUTOMATED EXTERNAL DEFIBRILLATOR (AED) OPERATION13–2 www.zoll.com 9650-0806-01 Rev. CIn AED operation, there is an Analysis/CPR Protocol mo

Page 72

AED Operation9650-0806-01 Rev. C Propaq MD Operator’s Guide 13–3Prepare PatientRemove all clothing covering the patient’s chest. Dry chest if necessar

Page 73 - Continuous Waveform Recording

CHAPTER 13 AUTOMATED EXTERNAL DEFIBRILLATOR (AED) OPERATION13–4 www.zoll.com 9650-0806-01 Rev. CSelect EnergyThe default energy selections for adult p

Page 74 - CHAPTER 4TRENDS

AED Operation9650-0806-01 Rev. C Propaq MD Operator’s Guide 13–52 AnalyzeWARNING! Do not analyze the patient ECG during patient movement. A patient mu

Page 75 - Chapter 5

CHAPTER 13 AUTOMATED EXTERNAL DEFIBRILLATOR (AED) OPERATION13–6 www.zoll.com 9650-0806-01 Rev. CWhen a nonshockable rhythm is detected, the unit displ

Page 76 - Alarm Indicator Self-Test

AED Operation9650-0806-01 Rev. C Propaq MD Operator’s Guide 13–73 Press SHOCKWARNING! Warn all persons in attendance of the patient to STAND CLEAR pri

Page 77 - Patient Alarm Display

CHAPTER 13 AUTOMATED EXTERNAL DEFIBRILLATOR (AED) OPERATION13–8 www.zoll.com 9650-0806-01 Rev. CContinue Patient CareContinue patient care according t

Page 78 - Equipment Alert Display

AED Operation9650-0806-01 Rev. C Propaq MD Operator’s Guide 13–9CHECK PULSEIf configured to do so, this message is displayed and announced in the foll

Page 79 - Suspending Alarms

Symbols Used on the Equipment9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-5Equipotentiality.Alternating current (ac).Direct current (dc).Auxiliary

Page 80

CHAPTER 13 AUTOMATED EXTERNAL DEFIBRILLATOR (AED) OPERATION13–10 www.zoll.com 9650-0806-01 Rev. CSwitching to Manual Mode OperationPress the Manual Mo

Page 81 - Alarm Options

9650-0806-01 Rev. C Propaq MD Operator’s Guide 14-1Chapter 1412-Lead ECG Interpretive AnalysisThe Propaq MD 12-Lead input is Type CF defibrillator pro

Page 82 - CHAPTER 5ALARMS

CHAPTER 14 12-LEAD ECG INTERPRETIVE ANALYSIS14-2 www.zoll.com 9650-0806-01 Rev. C• Wait 15 seconds after defibrillator discharge before attempting a

Page 83

Entering Patient Information9650-0806-01 Rev. C Propaq MD Operator’s Guide 14-3The Propaq MD unit uses the name that you enter in the Patient Info pan

Page 84

CHAPTER 14 12-LEAD ECG INTERPRETIVE ANALYSIS14-4 www.zoll.com 9650-0806-01 Rev. CEntering Patient Age and GenderThe Patient Info parameter panel prov

Page 85 - Monitoring ECG

12-Lead ECG Monitoring Setup9650-0806-01 Rev. C Propaq MD Operator’s Guide 14-5Applying Electrodes to the PatientDepending on local usage, ECG lead wi

Page 86 - ECG Monitoring Setup

CHAPTER 14 12-LEAD ECG INTERPRETIVE ANALYSIS14-6 www.zoll.com 9650-0806-01 Rev. CAvoid placing electrodes over tendons and major muscle masses.Make s

Page 87

12-Lead ECG Monitoring Setup9650-0806-01 Rev. C Propaq MD Operator’s Guide 14-7Locating the V1/C1 position (fourth intercostal space) is critically im

Page 88 - 5-Lead Electrode Placement

CHAPTER 14 12-LEAD ECG INTERPRETIVE ANALYSIS14-8 www.zoll.com 9650-0806-01 Rev. CObserving the 12-Lead Waveform TracesTo observe the 12-Lead waveform

Page 89

12-Lead ECG Monitoring Setup9650-0806-01 Rev. C Propaq MD Operator’s Guide 14-9After acquiring the ECG data, the unit saves the data and displays the

Page 90

CHAPTER 1GENERAL INFORMATION1-6 www.zoll.com 9650-0806-01 Rev. CDo not discard in trash. Recycle or dispose of properly.Return to a collection site i

Page 91

CHAPTER 14 12-LEAD ECG INTERPRETIVE ANALYSIS14-10 www.zoll.com 9650-0806-01 Rev. CAnalysis Page 2 displays the identifying information that has been

Page 92

12-Lead ECG Monitoring Setup9650-0806-01 Rev. C Propaq MD Operator’s Guide 14-11You can then view four 12-Lead snapshot pages by pressing to step th

Page 93 - ECG Monitoring and Pacemakers

CHAPTER 14 12-LEAD ECG INTERPRETIVE ANALYSIS14-12 www.zoll.com 9650-0806-01 Rev. CPrinting 12-Lead Waveform TracesOnce you have determined that the P

Page 94

12-Lead Print and Display Options9650-0806-01 Rev. C Propaq MD Operator’s Guide 14-13or a 3 x 4 Cabrera format:The Propaq MD unit stores a minimum of

Page 95

CHAPTER 14 12-LEAD ECG INTERPRETIVE ANALYSIS14-14 www.zoll.com 9650-0806-01 Rev. CSelecting 12-Lead AcquireWhen enabled, the unit automatically print

Page 96 - Respiration/Breath Rate Meter

12-Lead Print and Display Options9650-0806-01 Rev. C Propaq MD Operator’s Guide 14-15You can specify the following trace display ranges:Note: The disp

Page 97

CHAPTER 14 12-LEAD ECG INTERPRETIVE ANALYSIS14-16 www.zoll.com 9650-0806-01 Rev. C

Page 98 - Respiration Rate Alarm Limits

9650-0806-01 Rev. C Propaq MD Operator’s Guide 15–1Chapter 15Manual DefibrillationEmergency Defibrillation Procedure with PaddlesWarning! To avoid ris

Page 99 - ---) and heart rates

MANUAL DEFIBRILLATION15–2 www.zoll.com 9650-0806-01 Rev. CDetermine the Patient’s Condition Following Local Medical ProtocolsVerify the following:• Un

Page 100

Emergency Defibrillation Procedure with Paddles9650-0806-01 Rev. C Propaq MD Operator’s Guide 15–3Apply Paddles to ChestApply the paddles firmly to th

Page 101 - 2. Press

Conventions9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-7ConventionsThis guide uses the following conventions:Within text, the names and labels fo

Page 102 - Acknowledging LTA Alarms

MANUAL DEFIBRILLATION15–4 www.zoll.com 9650-0806-01 Rev. CCaution Changing the selected energy while the unit is charging or charged causes the defibr

Page 103 - RESP System Message

Emergency Defibrillation Procedure with Paddles9650-0806-01 Rev. C Propaq MD Operator’s Guide 15–53 Deliver ShockWarning! Warn all persons in attendan

Page 104

MANUAL DEFIBRILLATION15–6 www.zoll.com 9650-0806-01 Rev. CEmergency Defibrillation Procedure with Hands-Free Therapy ElectrodesDetermine the Patient’s

Page 105 - Chapter 8

Emergency Defibrillation Procedure with Hands-Free Therapy Electrodes9650-0806-01 Rev. C Propaq MD Operator’s Guide 15–7Note: If it is not possible to

Page 106 - How does NIBP Work?

MANUAL DEFIBRILLATION15–8 www.zoll.com 9650-0806-01 Rev. C2 Charge DefibrillatorPress the CHARGE button on the front panel.To increase or decrease the

Page 107 - The NIBP Numeric Display

Internal Paddles9650-0806-01 Rev. C Propaq MD Operator’s Guide 15–93 Deliver ShockWarning! Warn all persons in attendance of the patient to STAND CLEA

Page 108 - Selecting the NIBP Cuff

MANUAL DEFIBRILLATION15–10 www.zoll.com 9650-0806-01 Rev. CVerification Prior to UseBefore each use with the Propaq MD, verify the proper operation of

Page 109 - Connecting the NIBP Cuff

Synchronized Cardioversion9650-0806-01 Rev. C Propaq MD Operator’s Guide 15–11Synchronized CardioversionWarning! Only skilled personnel trained in Adv

Page 110

MANUAL DEFIBRILLATION15–12 www.zoll.com 9650-0806-01 Rev. CSynchronized Cardioversion ProcedureDetermine the Patient’s Condition and Provide Care Foll

Page 111

Synchronized Cardioversion Procedure9650-0806-01 Rev. C Propaq MD Operator’s Guide 15–13Unless otherwise configured, the unit automatically exits Sync

Page 112

Use this guide with MP App software version 02.06.07.00 or higher.The issue date for the Propaq MD Operator's Guide (REF 9650-0806-01 Rev. C) is

Page 113

CHAPTER 1GENERAL INFORMATION1-8 www.zoll.com 9650-0806-01 Rev. CWhen the pediatric patient is less than 8 years of age or weighs less than 55 lbs. (2

Page 114

MANUAL DEFIBRILLATION15–14 www.zoll.com 9650-0806-01 Rev. C3 Deliver ShockWarning! Warn all persons in attendance of the patient to STAND CLEAR prior

Page 115 - 200 250 300

9650-0806-01 Rev. C Propaq MD Operator’s Guide 16–1Chapter 16Advisory DefibrillationWhen the Propaq MD unit is configured for Advisory (or single anal

Page 116 - Specify NIBP Mode

CHAPTER 16 ADVISORY DEFIBRILLATION16–2 www.zoll.com 9650-0806-01 Rev. CAdvisory Defibrillation ProcedureDetermine the Patient’s Condition Following Lo

Page 117 - Start/Stop TurboCuf

Advisory Defibrillation Procedure9650-0806-01 Rev. C Propaq MD Operator’s Guide 16–31 Turn on unitPress the green power switch located on the top of t

Page 118 - NIBP System Messages

CHAPTER 16 ADVISORY DEFIBRILLATION16–4 www.zoll.com 9650-0806-01 Rev. C2 Press ANALYZEWARNING! Keep patient motionless during ECG analysis. Do not tou

Page 119

Advisory Defibrillation Procedure9650-0806-01 Rev. C Propaq MD Operator’s Guide 16–5Note: When a nonshockable rhythm is detected, the Propaq MD does n

Page 120

CHAPTER 16 ADVISORY DEFIBRILLATION16–6 www.zoll.com 9650-0806-01 Rev. CRepeat AnalysisPress the ANALYZE quick access key to restart an ECG analysis an

Page 121 - Monitoring CO

9650-0806-01 Rev. C Propaq MD Operator’s Guide 17–1Chapter 17Advisory/CPR Protocol DefibrillationWhen the Propaq MD unit is configured for Advisory/CP

Page 122 - Monitoring Setup and Use

CHAPTER 17 ADVISORY/CPR PROTOCOL DEFIBRILLATION17–2 www.zoll.com 9650-0806-01 Rev. CWARNING! Use only pediatric electrodes to defibrillate patients un

Page 123 - CO2 Monitoring Setup and Use

Advisory/CPR Protocol Defibrillation Procedure9650-0806-01 Rev. C Propaq MD Operator’s Guide 17–31 Turn on unitPress the green power switch located on

Page 124 - Sampling Lines

Propaq MD Indications for Use9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-9ECG MonitoringThe Propaq MD is intended for use to monitor and/or recor

Page 125 - Applying a FilterLine Set

CHAPTER 17 ADVISORY/CPR PROTOCOL DEFIBRILLATION17–4 www.zoll.com 9650-0806-01 Rev. C2 Press ANALYZEWARNING! Keep patient motionless during ECG analysi

Page 126 - CHAPTER 9MONITORING CO2

Advisory/CPR Protocol Defibrillation Procedure9650-0806-01 Rev. C Propaq MD Operator’s Guide 17–5When you press the Select Energy Up/Down arrow keys o

Page 127 - Measuring CO2

CHAPTER 17 ADVISORY/CPR PROTOCOL DEFIBRILLATION17–6 www.zoll.com 9650-0806-01 Rev. C

Page 128 - and Respiration Rate Alarms

9650-0806-01 Rev. C Propaq MD Operator’s Guide 18–1Chapter 18External PacingWARNING! Pacing is intended for use on adult patients and on adolescent, c

Page 129 - Setting Upper FiCO

CHAPTER 18 EXTERNAL PACING18–2 www.zoll.com 9650-0806-01 Rev. CExternal PacingPropaq MD defibrillators contain a pacemaker for the emergency treatment

Page 130

Pacing in Demand Mode9650-0806-01 Rev. C Propaq MD Operator’s Guide 18–32 Apply ECG Electrodes/Hands-Free Therapy ElectrodesApply ECG electrodes, atta

Page 131 - System Messages

CHAPTER 18 EXTERNAL PACING18–4 www.zoll.com 9650-0806-01 Rev. C4 Set ModeUse the arrow keys to navigate to Mode, press the Select button, and then use

Page 132 - NO IMPLIED LICENSE

Pacing in Demand Mode9650-0806-01 Rev. C Propaq MD Operator’s Guide 18–5WARNING! Determination of electrical capture should only be performed by viewi

Page 133 - , SPCO, and SpMet)

CHAPTER 18 EXTERNAL PACING18–6 www.zoll.com 9650-0806-01 Rev. CPacing in Fixed ModeIf ECG electrodes are not available or there is some circumstance t

Page 134 - , General

Pacing in Fixed Mode9650-0806-01 Rev. C Propaq MD Operator’s Guide 18–73 Press Pacer buttonPress the Pacer button on the front panel of the unit. The

Page 135 - , Oximeter Sensor

CHAPTER 1GENERAL INFORMATION1-10 www.zoll.com 9650-0806-01 Rev. Cmotion and motion conditions, and for patients who are well or poorly perfused, in h

Page 136 - 4. Configure SpO

CHAPTER 18 EXTERNAL PACING18–8 www.zoll.com 9650-0806-01 Rev. C5 mA increments when decreasing the output. Observe the ECG for evidence of electrical

Page 137 - Detector Window

Pacing in Fixed Mode9650-0806-01 Rev. C Propaq MD Operator’s Guide 18–9Pediatric PacingNoninvasive pacing of pediatric patients is performed in an ide

Page 138 - Clear Protective

CHAPTER 18 EXTERNAL PACING18–10 www.zoll.com 9650-0806-01 Rev. C

Page 139

9650-0806-01 Rev. C Propaq MD Operator’s Guide 19–1Chapter 19Real CPR HelpWARNING! The Real CPR Help function is not intended for use on patients unde

Page 140 - Displaying SpO

CHAPTER 19 REAL CPR HELP19–2 www.zoll.com 9650-0806-01 Rev. CCPR Voice PromptsThe Propaq MD unit can be configured to issue voice prompts related to t

Page 141 - Enabling/Disabling SpO

CPR Dashboard9650-0806-01 Rev. C Propaq MD Operator’s Guide 19–3Constant Metronome DisabledAED Mode• If Enable Metronome is set to Always or AED Mode

Page 142 - Parameter Control Panel

CHAPTER 19 REAL CPR HELP19–4 www.zoll.com 9650-0806-01 Rev. CNote: If no compressions have been detected for at least 3 seconds or if the compressions

Page 143 - System Messages

CPR Dashboard9650-0806-01 Rev. C Propaq MD Operator’s Guide 19–5CPR Countdown TimerThe Propaq MD unit displays a CPR Countdown Timer to indicate the t

Page 144 - System Message Cause

CHAPTER 19 REAL CPR HELP19–6 www.zoll.com 9650-0806-01 Rev. CCPR Compression Bar GraphThe Propaq MD unit can display a CPR compression bar graph next

Page 145 - Chapter 11

9650-0806-01 Rev. C Propaq MD Operator’s Guide 20–1Chapter 20See-Thru CPRSee-Thru CPR is an optional feature in the Propaq MD.WARNING! The See-Thru CP

Page 146 - IBP Setup

Propaq MD Product Functions9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-11Defibrillator Output EnergyPropaq MD defibrillators can deliver biphasic

Page 147 - Zeroing the Transducer

CHAPTER 20 SEE-THRU CPR20–2 www.zoll.com 9650-0806-01 Rev. Cthe ECG rhythm, requiring the rescuer to stop CPR to assess the ECG. For example, in the c

Page 148 - Rezeroing a Transducer

Using See-Thru CPR9650-0806-01 Rev. C Propaq MD Operator’s Guide 20–3The following figure shows a patient in Fine VF. It is difficult for a rescuer to

Page 149 - Displaying IBP Measurements

CHAPTER 20 SEE-THRU CPR20–4 www.zoll.com 9650-0806-01 Rev. CThe following figure shows a patient in VF, which, during compressions, is slightly more d

Page 150

Using See-Thru CPR9650-0806-01 Rev. C Propaq MD Operator’s Guide 20–5The following figure shows a patient in PEA, which could easily be mistaken for F

Page 151

CHAPTER 20 SEE-THRU CPR20–6 www.zoll.com 9650-0806-01 Rev. CThe following figure shows a patient with an organized rhythm where See-Thru CPR effective

Page 152 - Setting IBP Source Label

9650-0806-01 Rev. C Propaq MD Operator’s Guide 21–1Chapter 21Patient DataThis chapter describes procedures for storing, viewing, and transferring pati

Page 153 - IBP System Messages

CHAPTER 21 PATIENT DATA21–2 www.zoll.com 9650-0806-01 Rev. CNote: The Propaq MD unit retains stored cases even if you turn off or remove the auxiliary

Page 154

Treatment Summary Report9650-0806-01 Rev. C Propaq MD Operator’s Guide 21–3Note: If you attempt to print a snapshot before the unit has finished stori

Page 155 - Monitoring Temperature

CHAPTER 21 PATIENT DATA21–4 www.zoll.com 9650-0806-01 Rev. CPrinting Treatment Summary ReportTo print a treatment summary report, do the following:1.

Page 156 - Displaying Temperature

Transferring Data to a USB Device9650-0806-01 Rev. C Propaq MD Operator’s Guide 21–5Warning! To avoid a possible shock hazard, do NOT make any electri

Page 157

CHAPTER 1GENERAL INFORMATION1-12 www.zoll.com 9650-0806-01 Rev. CYou should always check the expiration date on the electrode packaging. Do not use e

Page 158

CHAPTER 21 PATIENT DATA21–6 www.zoll.com 9650-0806-01 Rev. C

Page 159 - Temperature System Messages

9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–1Chapter 22CommunicationsIf installed, the Propaq MD unit can send data through an Ethernet connecti

Page 160

CHAPTER 22 COMMUNICATIONS22–2 www.zoll.com 9650-0806-01 Rev. CThe Communications IconThe communications icon is located to the right of the patient mo

Page 161 - (AED) Operation

The Wireless Menu9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–3The Wireless MenuThe Wireless menu has the following options: WiFi Access Point, C

Page 162 - AED Operation

CHAPTER 22 COMMUNICATIONS22–4 www.zoll.com 9650-0806-01 Rev. CCreating a Temporary Access Point ProfileWhen the unit has been moved to a temporary loc

Page 163

Creating a Temporary Access Point Profile9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–5Profile NameThe profile name is “Temporary Profile” and ca

Page 164 - Select Energy

CHAPTER 22 COMMUNICATIONS22–6 www.zoll.com 9650-0806-01 Rev. CNon-broadcast SSIDUse the navigation keys to enable or disable non-broadcast or hidden S

Page 165 - 2 Analyze

Creating a Temporary Access Point Profile9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–7Use the alphabetical keypad to enter the security key. Pre

Page 166

CHAPTER 22 COMMUNICATIONS22–8 www.zoll.com 9650-0806-01 Rev. CNote: The Propaq MD unit only supports PEAPv0/v1 + MSCHAPV2.TLS (Transport Layer Securit

Page 167 - 3 Press SHOCK

Creating a Temporary Access Point Profile9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–9The unit will then scan for nearby Bluetooth devices that

Page 168 - Audio and Display Messages

Propaq MD Product Functions9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-13After you recalibrate the battery, the Recalibration LED will only flash

Page 169

CHAPTER 22 COMMUNICATIONS22–10 www.zoll.com 9650-0806-01 Rev. CChanging the PINIt may be necessary to change the PIN on the Propaq MD unit; check your

Page 170

Setting up Communications in the Supervisor Menu9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–113. Enter the password. The Supervisor menu appears

Page 171 - Chapter 14

CHAPTER 22 COMMUNICATIONS22–12 www.zoll.com 9650-0806-01 Rev. CBluetoothUse the navigation keys to enable or disable this function.When disabled, all

Page 172 - Entering Patient Information

Setting up Communications in the Supervisor Menu9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–13WiFi Access Point ProfilesTo select and modify WiF

Page 173

CHAPTER 22 COMMUNICATIONS22–14 www.zoll.com 9650-0806-01 Rev. CProfile NameUse the alphabetical keypad to enter the profile name. Press SAVE to save c

Page 174 - 12-Lead ECG Monitoring Setup

Setting up Communications in the Supervisor Menu9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–15SSIDUse the alphabetical keypad to enter the SSID

Page 175

CHAPTER 22 COMMUNICATIONS22–16 www.zoll.com 9650-0806-01 Rev. CFor WPA-PSK or WPA2-PSK, you are required to enter a security key (8-64 characters).Use

Page 176

Setting up Communications in the Supervisor Menu9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–17Authentication Protocol (TLS or PEAP). Some fields

Page 177

CHAPTER 22 COMMUNICATIONS22–18 www.zoll.com 9650-0806-01 Rev. CSelect Load from USB. A list of files on the USB device will appear.Filenames can only

Page 178 - 00:17:43

Setting up Communications in the Supervisor Menu9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–19Press to return to the Enterprise Wi-fi Configur

Page 179

CHAPTER 1GENERAL INFORMATION1-14 www.zoll.com 9650-0806-01 Rev. CSafety ConsiderationsAll operators should review these safety considerations before

Page 180

CHAPTER 22 COMMUNICATIONS22–20 www.zoll.com 9650-0806-01 Rev. CConfiguring Cell Phone Provider InformationUse the arrow keys to enter information abou

Page 181

Setting up Communications in the Supervisor Menu9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–21The unit will then scan for nearby Bluetooth devic

Page 182

CHAPTER 22 COMMUNICATIONS22–22 www.zoll.com 9650-0806-01 Rev. CYou can then edit the settings of the paired device, and give the device an optional ni

Page 183 - . Using the navigation keys

Setting up Communications in the Supervisor Menu9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–23Setting up an Ethernet Connection An Ethernet adap

Page 184 - Selecting 12-Lead Acquire

CHAPTER 22 COMMUNICATIONS22–24 www.zoll.com 9650-0806-01 Rev. CConfiguring Report TransmissionsThe Propaq MD unit can send 12-lead reports to the ZOLL

Page 185 - Enabling 12-Lead Analysis

Sending a 12-lead report9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–254. Press . A list of preconfigured distribution lists appears.Use the nav

Page 186

CHAPTER 22 COMMUNICATIONS22–26 www.zoll.com 9650-0806-01 Rev. CCommunications System MessagesThe Propaq MD unit may display one of the following statu

Page 187 - Manual Defibrillation

Transmission Status of 12 Lead Snapshots9650-0806-01 Rev. C Propaq MD Operator’s Guide 22–27Transmission Status of 12 Lead Snapshots When you press th

Page 188 - 1 Select Energy Level

CHAPTER 22 COMMUNICATIONS22–28 www.zoll.com 9650-0806-01 Rev. C

Page 189 - 2 Charge Defibrillator

9650-0806-01 Rev. C Propaq MD Operator’s Guide 23–1Chapter 23PrintingThis chapter describes procedures for printing permanent records of pertinent pat

Page 190 - 200 J120 J

Warnings9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-15The Propaq MD unit might not perform to specifications when stored at the upper or lower ex

Page 191 - 3 Deliver Shock

CHAPTER 23 PRINTING23–2 www.zoll.com 9650-0806-01 Rev. CPrinter SetupTo configure the printer functions, press the More quick access key ( ), press th

Page 192 - Therapy Electrodes

Printing Patient Data9650-0806-01 Rev. C Propaq MD Operator’s Guide 23–3The SpO2 PlethysmographThe SpO2 plethysmograph is printed with major divisions

Page 193

CHAPTER 23 PRINTING23–4 www.zoll.com 9650-0806-01 Rev. CPrinting TrendsPrinted trends are useful for reviewing the patient’s vital signs over the last

Page 194

9650-0806-01 Rev. C Propaq MD Operator’s Guide 24-1Chapter 24MaintenanceResuscitation equipment must be maintained to be ready for immediate use. To e

Page 195 - Internal Paddles

CHAPTER 24 MAINTENANCE24-2 www.zoll.com 9650-0806-01 Rev. CDaily/Shift Check ProcedureInspectionEquipment and Accessories• Ensure that the Propaq MD

Page 196 - Verification Prior to Use

Defibrillator/Pacing Test with Hands-Free Therapy Electrodes9650-0806-01 Rev. C Propaq MD Operator’s Guide 24-3Defibrillator/Pacing Test with Hands-Fr

Page 197 - Synchronized Cardioversion

CHAPTER 24 MAINTENANCE24-4 www.zoll.com 9650-0806-01 Rev. C8 Press the Select Energy arrow (up or down). Verify that the defibrillation window displa

Page 198 - Press the Sync Button

Defibrillator Testing with External Paddles9650-0806-01 Rev. C Propaq MD Operator’s Guide 24-5Defibrillator Testing with External PaddlesPrior to test

Page 199

CHAPTER 24 MAINTENANCE24-6 www.zoll.com 9650-0806-01 Rev. C12 While the defibrillator is still charged, press the paddles electrodes together then pr

Page 200

Recommended Minimum Preventive Maintenance Schedule9650-0806-01 Rev. C Propaq MD Operator’s Guide 24-7Recommended Minimum Preventive Maintenance Sched

Page 201 - Advisory Defibrillation

CHAPTER 1GENERAL INFORMATION1-16 www.zoll.com 9650-0806-01 Rev. CEquipment such as electrocautery or diathermy equipment, RFID readers, electronic ar

Page 202

CHAPTER 24 MAINTENANCE24-8 www.zoll.com 9650-0806-01 Rev. CCleaning instructionsCleaning the Propaq MD unitTo clean the Propaq MD unit, use a nearly

Page 203 - 1 Turn on unit

Cleaning instructions9650-0806-01 Rev. C Propaq MD Operator’s Guide 24-9Cleaning SpO2 SensorsClean the SpO2 sensors with a cloth that has been slightl

Page 204 - 2 Press ANALYZE

CHAPTER 24 MAINTENANCE24-10 www.zoll.com 9650-0806-01 Rev. C2. Refer to the illustration inside of the paper compartment for proper paper orientation

Page 205 - 3 Press SHOCK button

9650-0806-01 Rev. C Propaq MD Operator’s Guide 24-11Signature _______________________________________________Propaq MD OPERATOR’S SHIFT CHECKLISTDate:

Page 206 - Continue Patient Care

CHAPTER 24 MAINTENANCE24-12 www.zoll.com 9650-0806-01 Rev. C

Page 207 - Chapter 17

9650-0806-01 Rev. C Propaq MD Operator’s Guide A-1Appendix ASpecificationsThis chapter provides specification information for the Propaq MD Monitor/De

Page 208

SpecificationsA-2 www.zoll.com 9650-0806-01 Rev. CDefibrillatorCharge Time:• Less than 7 seconds with a new, fully charged battery (first 15 charges

Page 209

Defibrillator9650-0806-01 Rev. C Propaq MD Operator’s Guide A-3* For all energy levels, accuracy is equal to either ±15% or 3 joules, whichever is gre

Page 210

SpecificationsA-4 www.zoll.com 9650-0806-01 Rev. CThe vertical axis shows the current in amperes (A); the horizontal axis shows the duration of time

Page 211

Defibrillator9650-0806-01 Rev. C Propaq MD Operator’s Guide A-5Figure A-3. Rectilinear Biphasic Waveform at 120 JoulesFigure A-4. Rectilinear Biphasic

Page 212

Warnings9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-17Before charging the defibrillator, verify that the energy selected on the display is the de

Page 213 - External Pacing

SpecificationsA-6 www.zoll.com 9650-0806-01 Rev. CFigure A-5. Rectilinear Biphasic Waveform at 85 JoulesFigure A-6. Rectilinear Biphasic Waveform at

Page 214 - Pacing in Demand Mode

Defibrillator9650-0806-01 Rev. C Propaq MD Operator’s Guide A-7Figure A-7. Rectilinear Biphasic Waveform at 50 JoulesFigure A-8. Rectilinear Biphasic

Page 215

SpecificationsA-8 www.zoll.com 9650-0806-01 Rev. CFigure A-9. Rectilinear Biphasic Waveform at 20 JoulesFigure A-10. Rectilinear Biphasic Waveform at

Page 216 - 8 Determine Capture

Defibrillator9650-0806-01 Rev. C Propaq MD Operator’s Guide A-9Figure A-11. Rectilinear Biphasic Waveform at 10 JoulesFigure A-12. Rectilinear Biphasi

Page 217

SpecificationsA-10 www.zoll.com 9650-0806-01 Rev. CFigure A-13. Rectilinear Biphasic Waveform at 8 JoulesFigure A-14. Rectilinear Biphasic Waveform a

Page 218 - Pacing in Fixed Mode

Defibrillator9650-0806-01 Rev. C Propaq MD Operator’s Guide A-11Figure A-15. Rectilinear Biphasic Waveform at 6 JoulesFigure A-16. Rectilinear Biphasi

Page 219

SpecificationsA-12 www.zoll.com 9650-0806-01 Rev. CFigure A-17. Rectilinear Biphasic Waveform at 4 JoulesFigure A-18. Rectilinear Biphasic Waveform a

Page 220 - 9 Determine Optimum Threshold

Defibrillator9650-0806-01 Rev. C Propaq MD Operator’s Guide A-13Figure A-19. Rectilinear Biphasic Waveform at 2 JoulesFigure A-20. Rectilinear Biphasi

Page 221

SpecificationsA-14 www.zoll.com 9650-0806-01 Rev. CMonitor/DisplayInput: 3-lead, 5-lead, or 12-lead patient cable, paddles, Multifunction or OneStep

Page 222 - CHAPTER 18 EXTERNAL PACING

Impedance Pneumography9650-0806-01 Rev. C Propaq MD Operator’s Guide A-15Heart Rate Response to Irregular Rhythm: (AAMI EC13-2002, section 4.1.2.1.e.)

Page 223 - Real CPR Help

9650-0806-01 Rev. C Propaq MD Operator’s Guide iTable of ContentsChapter 1 General InformationProduct Description ...

Page 224 - CPR Metronome

CHAPTER 1GENERAL INFORMATION1-18 www.zoll.com 9650-0806-01 Rev. CTranscutaneous pacing may cause discomfort ranging from mild to severe, depending on

Page 225 - CPR Dashboard

SpecificationsA-16 www.zoll.com 9650-0806-01 Rev. CAlarmsHeart Rate Alarms:Audible: 5 pulse, 900 Hz tone, with a PW of 125 msec, a PRI of 250 msec, a

Page 226 - Chest Compression Indicator

Recorder9650-0806-01 Rev. C Propaq MD Operator’s Guide A-17RecorderType: High-resolution thermal array.Annotation: Time, date, ECG lead, ECG gain, hea

Page 227

SpecificationsA-18 www.zoll.com 9650-0806-01 Rev. CGeneralWeight: 10.6 lbs.without battery and paper11.7 lbs.with battery and paperDimensions: Withou

Page 228 - CPR Compression Bar Graph

Pacer9650-0806-01 Rev. C Propaq MD Operator’s Guide A-19PacerType: External transcutaneous pacingPacer Rate: 30 to 180 BPM ± 1.5%.Output Current: 0 to

Page 229 - Chapter 20

SpecificationsA-20 www.zoll.com 9650-0806-01 Rev. CPulse OximeterRange: Oxygen Saturation (% SpO2) 1% - 100%Carboxyhemoglobin Saturation (% SpCO) 0

Page 230 - Using See-Thru CPR

Pulse Oximeter9650-0806-01 Rev. C Propaq MD Operator’s Guide A-21Energies (Radiant Power) of light for Rainbow Sensors at 100 mA pulsed: 25 mWBio-Com

Page 231

SpecificationsA-22 www.zoll.com 9650-0806-01 Rev. CNon-Invasive Blood PressureTechnique: Non-invasive oscillometric methodOperating Modes: Automatic

Page 232 - CHAPTER 20 SEE-THRU CPR

Invasive Pressures9650-0806-01 Rev. C Propaq MD Operator’s Guide A-23Typical Determination time without Artifact:Measurements on the deflation: 30 to

Page 233

SpecificationsA-24 www.zoll.com 9650-0806-01 Rev. CTemperatureNumber of Channels: 2Measurement Range: 0° to 50° CAccuracy: ± 0.1° C from 10° C to 50°

Page 234

Clinical Trial Results for the Biphasic Waveform9650-0806-01 Rev. C Propaq MD Operator’s Guide A-25Clinical Trial Results for the Biphasic WaveformThe

Page 235 - Patient Data

Warnings9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-19In some instances, such as obstructed airway, the patient's breathing attempts may not

Page 236 - The log is empty

SpecificationsA-26 www.zoll.com 9650-0806-01 Rev. CSuccessful defibrillation with rectilinear biphasic shocks was achieved with 58% less delivered cu

Page 237 - Treatment Summary Report

Clinical Trial Results for the Biphasic Waveform9650-0806-01 Rev. C Propaq MD Operator’s Guide A-27Results: The study population of 165 patients had a

Page 238 - USB Device Port

SpecificationsA-28 www.zoll.com 9650-0806-01 Rev. CSynchronized Cardioversion of Atrial FibrillationCardioversion of atrial fibrillation (AF) and ove

Page 239 - Clearing the Log

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration9650-0806-01 Rev. C Propaq MD Operator’s Guide A-29Electromagnetic Compatibility

Page 240 - CHAPTER 21 PATIENT DATA

SpecificationsA-30 www.zoll.com 9650-0806-01 Rev. CElectromagnetic Immunity (IEC 60601-1-2)Guidance and manufacturer’s declaration – electromagnetic

Page 241 - Communications

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration9650-0806-01 Rev. C Propaq MD Operator’s Guide A-31Electromagnetic Immunity: Life

Page 242 - The Communications Icon

SpecificationsA-32 www.zoll.com 9650-0806-01 Rev. CElectromagnetic Immunity: Non Life-Supporting FunctionsNon life-supporting functions of the Propaq

Page 243 - The Wireless Menu

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration9650-0806-01 Rev. C Propaq MD Operator’s Guide A-33Recommended Separation Distanc

Page 244 - CHAPTER 22 COMMUNICATIONS

SpecificationsA-34 www.zoll.com 9650-0806-01 Rev. CRecommended Separation Distances from RF Equipment for the Propaq MD IBP Non Life-Supporting Funct

Page 245 - Network Setting

ECG Analysis Algorithm Accuracy9650-0806-01 Rev. C Propaq MD Operator’s Guide A-35ECG Analysis Algorithm AccuracySensitivity and specificity are expre

Page 246 - Authentication

CHAPTER 1GENERAL INFORMATION1-20 www.zoll.com 9650-0806-01 Rev. CCO2During MRI scanning, the monitor must be placed outside the MRI suite. When the m

Page 247

SpecificationsA-36 www.zoll.com 9650-0806-01 Rev. CWireless Output Guidance and Manufacturer’s DeclarationRF Transmission Emitted (IEC 60601-1-2)The

Page 248 - Bluetooth Device Pairing

9650-0806-01 Rev. C Propaq MD Operator’s Guide B-1Appendix BAccessoriesThe following accessories are compatible for use with the Propaq MD unit. To or

Page 249

ACCESSORIESB-2 www.zoll.com 9650-0806-01 Rev. CFilterLine H Set, Adult/Pediatric, box of 25FilterLine H Set, Infant/Neonate, box of 25FilterLine Set,

Page 250 - Changing the PIN

9650-0806-01 Rev. C Propaq MD Operator’s Guide B-3Rainbow DCI-dc8, 8' Adult Reusable Patient Cable / SensorRainbow DCI-dc12, 12' Adult Reusa

Page 251 - Wireless

ACCESSORIESB-4 www.zoll.com 9650-0806-01 Rev. CCPR stat-padz HVP Multifunction CPR Electrodes - 8 pair/case CPR stat-padz HVP Multifunction CPR Elect

Page 252 - HAPTER 22 COMMUNICATIONS

9650-0806-01 Rev. C Propaq MD Operator’s Guide B-5Power AccessoriesZOLL SurePower II Rechargable BatterySurePower Charger StationSurePower II Battery

Page 253 - WiFi Access Point Profiles

ACCESSORIESB-6 www.zoll.com 9650-0806-01 Rev. C

Page 254

Warnings9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-21If the LOW BATTERY icon appears, plug the Propaq MD unit into a power source or install a f

Page 255 - Authenication Protocol

CHAPTER 1GENERAL INFORMATION1-22 www.zoll.com 9650-0806-01 Rev. CDo not touch the bed, patient, or any equipment connected to the patient during defi

Page 256

Cautions9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-23During electrosurgery, observe the following guidelines to minimize electrosurgery unit (ES

Page 257

CHAPTER 1GENERAL INFORMATION1-24 www.zoll.com 9650-0806-01 Rev. CRestarting the DefibrillatorCertain events require the Propaq MD products to be rest

Page 258

Software License9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-25Notification of Adverse EventsAs a health care provider, you may have responsibilit

Page 259

CHAPTER 1GENERAL INFORMATION1-26 www.zoll.com 9650-0806-01 Rev. CServiceThe Propaq MD only requires recalibration of the CO2 module. Service is requi

Page 260

The ZOLL Serial Number9650-0806-01 Rev. C Propaq MD Operator’s Guide 1-27The ZOLL Serial NumberEach ZOLL product displays a serial number that contain

Page 261

TABLE OF CONTENTSii www.zoll.com 9650-0806-01 Rev. CWarnings...

Page 262

CHAPTER 1GENERAL INFORMATION1-28 www.zoll.com 9650-0806-01 Rev. C

Page 263

9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–1Chapter 2Product OverviewDefibrillator Controls and Indicators21578355694

Page 264 - Sending a 12-lead report

CHAPTER 2PRODUCT OVERVIEW2–2 www.zoll.com 9650-0806-01 Rev. CThe Front PanelThe front panel of the Propaq MD device includes the display screen, quick

Page 265

Defibrillator Controls and Indicators9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–3Table 2-2. Propaq MD Controls and IndicatorsControl or Indicato

Page 266

CHAPTER 2PRODUCT OVERVIEW2–4 www.zoll.com 9650-0806-01 Rev. CDisplay ScreenThe front panel includes a color display which shows:• Date and time• Patie

Page 267

Defibrillator Controls and Indicators9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–5• EtCO2 numeric data• Invasive pressure numeric data• The selec

Page 268

CHAPTER 2PRODUCT OVERVIEW2–6 www.zoll.com 9650-0806-01 Rev. CBattery Status and Auxiliary Power IndicatorsThe battery status indicator displays variou

Page 269 - Printing

Defibrillator Controls and Indicators9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–7Patient Cables and ConnectorsThe left and right sides of the un

Page 270 - Printing Waveforms

CHAPTER 2PRODUCT OVERVIEW2–8 www.zoll.com 9650-0806-01 Rev. CMultifunction Cable (MFC)The unit ships with an MFC that is used to defibrillate the pati

Page 271 - Printing Patient Data

Defibrillator Controls and Indicators9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–9Removing MFC from UnitTwist the connector to the left to unlock

Page 272 - Printing Trends

9650-0806-01 Rev. C Propaq MD Operator’s Guide iiiChapter 3 Monitoring OverviewPropaq MD Monitoring Functions ...

Page 273 - Maintenance

CHAPTER 2PRODUCT OVERVIEW2–10 www.zoll.com 9650-0806-01 Rev. CWhen disconnecting the OneStep electrode and OneStep cable, press down the latch with yo

Page 274 - Daily/Shift Check Procedure

Defibrillator Controls and Indicators9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–11If you need to detach the MFC from the APEX paddles, push the

Page 275 - Electrodes

CHAPTER 2PRODUCT OVERVIEW2–12 www.zoll.com 9650-0806-01 Rev. CFigure 2-10. OneStep Cable Connected to APEX PaddleIf you need to detach the OneStep ca

Page 276 - CHAPTER 24 MAINTENANCE

Defibrillator Controls and Indicators9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–13Pediatric-size electrodes are built in to the paddle assembly

Page 277

CHAPTER 2PRODUCT OVERVIEW2–14 www.zoll.com 9650-0806-01 Rev. CTo connect the auxiliary power adapter, align the white arrow of the auxiliary power con

Page 278

Navigating the Display Screen9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–15Table 2-3. Propaq MD Quick Access KeysQuick access key DescriptionLead

Page 279 - Annually

CHAPTER 2PRODUCT OVERVIEW2–16 www.zoll.com 9650-0806-01 Rev. CPrint Trends Prints the trends that are displayed in the Trend Summary window.Trend Sett

Page 280 - Cleaning instructions

Common Tasks9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–17Navigation Keys Use the navigation keys (up/clockwise arrow, down/counterclockwise arro

Page 281

CHAPTER 2PRODUCT OVERVIEW2–18 www.zoll.com 9650-0806-01 Rev. CChanging the Display BrightnessThe following procedure shows how to select the different

Page 282 - Cleaning the Print Head

Common Tasks9650-0806-01 Rev. C Propaq MD Operator’s Guide 2–19Using Treatment ButtonsPressing the Treatment quick access key ( ) causes the unit to d

Page 283

TABLE OF CONTENTSiv www.zoll.com 9650-0806-01 Rev. CChapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)Respiration/Breath Rate Meter ...

Page 284

CHAPTER 2PRODUCT OVERVIEW2–20 www.zoll.com 9650-0806-01 Rev. C

Page 285 - Specifications

9650-0806-01 Rev. C Propaq MD Operator’s Guide 3-1Chapter 3Monitoring OverviewThis chapter provides an overview of the Propaq MD unit’s monitoring fun

Page 286 - Defibrillator

CHAPTER 3MONITORING OVERVIEW3-2 www.zoll.com 9650-0806-01 Rev. CECG An ECG waveform trace appears at the top of the display area. You can specify tha

Page 287

Monitoring Display Options9650-0806-01 Rev. C Propaq MD Operator’s Guide 3-3Capnography (CO2) CO2 monitoring measures the CO2 concentration in a patie

Page 288

CHAPTER 3MONITORING OVERVIEW3-4 www.zoll.com 9650-0806-01 Rev. CYou can display up to four waveform traces that you specify on the Waveform Display w

Page 289

Monitoring Display Options9650-0806-01 Rev. C Propaq MD Operator’s Guide 3-5 06/06/2012 12:34:56Adult00:17:43I, II,III...12CO2AnalyzeIBP1 cm/mV

Page 290

CHAPTER 3MONITORING OVERVIEW3-6 www.zoll.com 9650-0806-01 Rev. CPress the Home/Display button when viewing the Trends Status window and the Large Num

Page 291

Configuring the Waveform Display9650-0806-01 Rev. C Propaq MD Operator’s Guide 3-7Configuring the Waveform DisplayYou can display up to four waveform

Page 292

CHAPTER 3MONITORING OVERVIEW3-8 www.zoll.com 9650-0806-01 Rev. CWhen the unit cascades the ECG Lead lI trace, the Waveform Display window appears as

Page 293

Configuring the Waveform Display9650-0806-01 Rev. C Propaq MD Operator’s Guide 3-9Inserting a third waveform trace for ECG lead aVR:06/06/2012

Page 294

9650-0806-01 Rev. C Propaq MD Operator’s Guide vChapter 10 Pulse CO-Oximetry (SpO2, SPCO, and SpMet)Warnings -- SpO2, General...

Page 295

CHAPTER 3MONITORING OVERVIEW3-10 www.zoll.com 9650-0806-01 Rev. CInserting a capnogram (CO2) into the fourth trace area: 06/06/2012 12:34:56Ad

Page 296

9650-0806-01 Rev. C Propaq MD Operator’s Guide 4-1Chapter 4TrendsThe Propaq MD unit accumulates a patient’s trend information by logging all monitored

Page 297

CHAPTER 4TRENDS4-2 www.zoll.com 9650-0806-01 Rev. CFigure 4-1 Trends Status windowThe Trends status window displays the logged trend information and

Page 298 - Monitor/Display

Changing the Trends Status Window Display9650-0806-01 Rev. C Propaq MD Operator’s Guide 4-3Changing the Trends Status Window DisplayBy default, the Tr

Page 299 - Impedance Pneumography

CHAPTER 4TRENDS4-4 www.zoll.com 9650-0806-01 Rev. C

Page 300 - • if source is IBP, 5 seconds

9650-0806-01 Rev. C Propaq MD Operator’s Guide 5-1Chapter 5AlarmsThe Propaq MD unit supports the detection and indication of patient alarms and techni

Page 301 - Recorder

CHAPTER 5ALARMS5-2 www.zoll.com 9650-0806-01 Rev. CAudible Alarm IndicatorsThe Propaq MD unit sounds an audible alarm to indicate the priority level

Page 302 - • EN 1789 for ambulance

Patient Alarm Display9650-0806-01 Rev. C Propaq MD Operator’s Guide 5-3Patient Alarm DisplayWhen a patient’s vital signs measurements trigger an alarm

Page 303 - Accuracy CO

CHAPTER 5ALARMS5-4 www.zoll.com 9650-0806-01 Rev. CLife Threatening Rhythm AlarmsWhen LTA monitoring is enabled, the Propaq MD will monitor for the f

Page 304

Responding to Active Alarms -- Silencing the Alarm9650-0806-01 Rev. C Propaq MD Operator’s Guide 5-5Responding to Active Alarms -- Silencing the Alarm

Page 305

TABLE OF CONTENTSvi www.zoll.com 9650-0806-01 Rev. CDisplaying Temperature...

Page 306 - Non-Invasive Blood Pressure

CHAPTER 5ALARMS5-6 www.zoll.com 9650-0806-01 Rev. CThe Alarm Suspension TimerDuring an alarm suspension, the window displays an alarm suspension time

Page 307 - A B C D E

Alarm Options9650-0806-01 Rev. C Propaq MD Operator’s Guide 5-7Alarm OptionsThe Propaq MD unit provides alarm options that you can specify through the

Page 308 - Temperature

CHAPTER 5ALARMS5-8 www.zoll.com 9650-0806-01 Rev. CSelecting Default Alarm LimitsThe first three options -- Default Adult, Default Pediatric, Default

Page 309 - First shock efficacy 93% 99%

Alarm Options9650-0806-01 Rev. C Propaq MD Operator’s Guide 5-9SpO2(%)Entire range Limit = 100 (Adult and Pediatric)Limit = Numeric + 5 (Neo-nate)Limi

Page 310 - Atrial Fibrillation (AF)

CHAPTER 5ALARMS5-10 www.zoll.com 9650-0806-01 Rev. C

Page 311

9650-0806-01 Rev. C Propaq MD Operator’s Guide 6-1Chapter 6Monitoring ECGThis chapter describes how to use the Propaq MD unit to monitor ECG.Propaq MD

Page 312 - Posterior

CHAPTER 6MONITORING ECG6-2 www.zoll.com 9650-0806-01 Rev. C• To avoid a shock hazard and interference from nearby electrical equipment, keep electrod

Page 313 - Manufacturer’s Declaration

ECG Monitoring Setup9650-0806-01 Rev. C Propaq MD Operator’s Guide 6-3Applying Electrodes to the PatientThe following sections show where to place ele

Page 314

CHAPTER 6MONITORING ECG6-4 www.zoll.com 9650-0806-01 Rev. C5-Lead Electrode PlacementDepending upon local usage, the ECG leads are marked either RA,

Page 315

ECG Monitoring Setup9650-0806-01 Rev. C Propaq MD Operator’s Guide 6-5Figure 6-2 5-Lead Electrode PlacementConnecting the ECG Cable To the Propaq MD U

Page 316 - : 12-lead ECG, EtCO

9650-0806-01 Rev. C Propaq MD Operator’s Guide viiChapter 15 Manual DefibrillationEmergency Defibrillation Procedure with Paddles...

Page 317 - Life-Supporting Functions

CHAPTER 6MONITORING ECG6-6 www.zoll.com 9650-0806-01 Rev. CSelecting ECG Waveforms for DisplayYou can fit up to four waveforms on the Propaq MD displ

Page 318

ECG Monitoring Setup9650-0806-01 Rev. C Propaq MD Operator’s Guide 6-7Select a current waveform source, and the unit displays the waveform immediately

Page 319 - References:

CHAPTER 6MONITORING ECG6-8 www.zoll.com 9650-0806-01 Rev. CSelecting the Waveform Trace SizeThe Propaq MD unit allows you to select the waveform trac

Page 320 - FCC Notice

ECG Monitoring and Pacemakers9650-0806-01 Rev. C Propaq MD Operator’s Guide 6-9ECG Monitoring and PacemakersWhen the unit performs ECG monitoring on a

Page 321 - Accessories

CHAPTER 6MONITORING ECG6-10 www.zoll.com 9650-0806-01 Rev. C

Page 322 - ACCESSORIES

9650-0806-01 Rev. C Propaq MD Operator’s Guide 7-1Chapter 7Monitoring Respiration (Resp) and Heart Rate (HR)This chapter describes how to use the Prop

Page 323

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)7-2 www.zoll.com 9650-0806-01 Rev. CNote: The respiration rate is disabled during defibril

Page 324

Configuring Respiration (RR/BR) Alarms and Settings9650-0806-01 Rev. C Propaq MD Operator’s Guide 7-3Configuring Respiration (RR/BR) Alarms and Settin

Page 325

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)7-4 www.zoll.com 9650-0806-01 Rev. CRespiration Rate Alarm LimitsInitially, the Resp Alarm

Page 326

Heart Rate Meter9650-0806-01 Rev. C Propaq MD Operator’s Guide 7-5Enabling/Disabling Resp MonitoringSelect the Resp Monitoring prompt to enable or dis