6 Fully Automatic AED Plus Administrator’s Guide
and START CPR, and illuminates the CPR-related graphics. A period of 2 minutes (depending upon
device configuration) is then allowed for rescuer CPR. Following this “CPR period”, the Fully
Automatic AED Plus automatically re-initiates a new ECG rhythm analysis.
The Fully Automatic AED Plus automatically adjusts defibrillation energy to adult or pediatric levels
based on the type of electrodes attached to the unit. In its factory default configuration, the unit delivers
the first three shocks at 120J, 150J, and 200J respectively in adult mode, and at 50J, 70J, and 85J in
pediatric mode. However, the device can be configured to deliver shocks at other energy levels, as long
as the setting for each successive shock is the same or greater than the setting for the preceding shock.
Shocks are delivered automatically; the heart icon (and associated LED) located in the center of the
graphical user interface will light up, and the unit will emit a loud tone when a shock is delivered.
Consult the Fully Automatic AED Plus Operators Guide for more details on the graphics included in
the unit’s graphic interface, the audio prompts issued at each step in the treatment protocol and the
rescuer action associated with these audio and visual prompts.
Loss of contact between the electrodes and the victim interrupts ECG analysis and/or shock delivery
until the electrodes are re-attached and results in a CHECK ELECTRODE PADS prompt to the operator.
Voice Prompts
During clinical use of the Fully Automatic AED Plus, you may hear the following voice prompts.
Table 2: Clinical Voice Prompts
Voice Prompt Definition
UNIT OK. Fully Automatic AED Plus has successfully passed its power up
self tests.
UNIT FAILED. Fully Automatic AED Plus has failed its power up self tests and
is not usable for victim care.
CHANGE BATTERIES. Fully Automatic AED Plus self test has detected a low battery
condition that is insufficient for the device’s use for victim care.
Replace batteries immediately.
STAY CALM. Relax as much as possible and focus on the rescue effort.
CHECK RESPONSIVENESS. Check victim for responsiveness/consciousness by gently
shaking the victim and shouting “Are you all right?”.
CALL FOR HELP. Activate the EMS system or ask a bystander to do it for you.
OPEN AIRWAY. Place victim in the supine position and perform Head Tilt – Chin
Lift or Jaw – Thrust maneuver to open victim’s airway. (This
prompt is off by default.)
CHECK BREATHING. Look, listen or feel for the presence of breathing and/or airflow
from the victim’s lungs. (This prompt is off by default.)
GIVE TWO BREATHS. If victim is not breathing, give two rescue breaths. (This prompt
is off by default.)
PLUG IN CABLE. Ensure that the electrode cable is properly connected to the Fully
Automatic AED Plus electrode connector.