24 Fully Automatic AED Plus Administrator’s Guide
Troubleshooting
The following table summarizes common error indications on the Fully Automatic AED Plus, and their
associated corrective action. Return the Fully Automatic AED Plus to ZOLLs Technical Service
Department if the Fully Automatic AED Plus is not working properly.
Table 5: Troubleshooting
Technical Problem Recommended Action
Self test failed. Initiate manual self test by pressing and holding the
ON/OFF Button for more than 5 seconds. Attempt
to repair the device by replacing the batteries or
electrodes. If the Fully Automatic AED Plus fails
test again, remove the Fully Automatic AED Plus
from service and contact ZOLL Technical Service.
CHANGE BATTERIES prompt. Replace all batteries with new batteries at the same
time. Press the Battery Reset Button when
prompted.
Red “X” in status indicator window. Initiate manual self test by pressing and holding the
ON/OFF Button for more than 5 seconds.
Check to see if cable is attached properly to the
Fully Automatic AED Plus or replace the
electrodes.
Cycle power on the Fully Automatic AED Plus by
turning the unit off, then on again.
Replace all batteries at the same time with new
batteries that are less than 1 year old. Press the
Battery Reset Button when prompted.
If the Fully Automatic AED Plus still does not
operate correctly, remove unit from service and
contact ZOLL Technical Service.
Beeping noise when Fully Automatic
AED Plus is off.
Remove Fully Automatic AED Plus from service
and replace batteries. Replace all batteries at the
same time with new batteries. Press the Battery
Reset Button when prompted.
If beeping continues, contact ZOLL Technical
Service.
PLUG IN CABLE prompt. Check cable connection between electrodes and the
Fully Automatic AED Plus.
ANALYSIS HALTED. KEEP VICTIM
STILL prompt.
Excessive artifact detected during ECG analysis.
Victim must be motionless during ECG analysis. Do
not touch the victim during analysis. Keep the
victim still. If the rescuer is using the Fully
Automatic AED Plus in an emergency vehicle,
bring the vehicle to a halt before performing ECG
analysis.