Fully Automatic AED Plus Administrator’s Guide 9
Using the LCD Display
The Fully Automatic AED Plus is equipped with a 1.3 x 2.6 inch LCD screen (see Figure 3) that
displays the following information:
Figure 3: LCD Displays
NOTE Some special Fully Automatic AED Plus models do not have an LCD.
Shock Count (Upper left corner of screen): Indicates the total number of defibrillation shocks
delivered by the Fully Automatic AED Plus since it was last powered on. Shock count is saved through
brief power off periods (< 5 seconds). When the Fully Automatic AED Plus is turned off for more than
5 seconds, the shock count is reset to 0.
Elapsed Time (Upper right corner of screen): Indicates the total time in minutes and seconds that has
elapsed since the Fully Automatic AED Plus was last powered on. Elapsed time continues to be
counted through brief power off periods (< 5 seconds). When the Fully Automatic AED Plus is turned
off for more than 5 seconds, elapsed time is reset to 00:00. When elapsed time exceeds 99 minutes and
59 seconds, the timer wraps around to 00:00 and continues counting.
CPR Depth of Compression Indicator (Right side of screen): A bar graph is displayed that shows
the depth of chest compressions measured during the delivery of CPR. Indicator lines are displayed in
the bar graph area at 2 and 2.4 inches of compression depth to provide reference points for rescuers
performing CPR.
Visual User Prompts (Lower 1/3 of screen): Whenever the Fully Automatic AED Plus issues a voice
prompt, the text of the voice prompt is simultaneously displayed on the LCD display.
ECG Waveform (Center portion of screen): Although Fully Automatic AED Plus units do not display
ECG waveforms in their factory default configuration, the device can be specifically set up to
continuously display ECG signals as they are acquired.
05:25
SHOCKS: 05
CHECK
RESPONSIVENESS
05:25
SHOCKS: 05
CHECK
RESPONSIVENESS
With ECG
Without ECG
2” (5 cm)
depth
2.4” (6 cm)
depth
Elapsed Time
CPR Depth of
Compression Indicator
2” (5 cm)
depth
2.4” (6 cm)
depth
Shock Count