Non-Interpretive12-Lead ECG Monitoring 9650-0218-01 Rev. B
OPTION INSERT 9650-0218-01 Non-Interpretive 12-lead - 8 12-Lead Summary Reports Immediately after acquisition, the 12-lead data prints in the followin
Non-Interpretive 12-lead 9650-0218-01 Non-Interpretive 12-lead - 9 2x6 Format (Fax Report Only) The "2x6 Format, 1 Page", prints all 12-
OPTION INSERT 9650-0218-01 Non-Interpretive 12-lead - 10 Patient Information DEVICE ID: MEDIC_1 RECORDED: 11:33:01 15 DEC 99 PATIENT NAME : JOE S
Non-Interpretive 12-lead 9650-0218-01 Non-Interpretive 12-lead - 11 Daily Operational Verification 1. Connect the V-lead cable to the 12-lead c
OPTION INSERT 9650-0218-01 Non-Interpretive 12-lead - 12 Troubleshooting The troubleshooting guide is intended to help you identify and correct proble
Non-Interpretive 12-lead9650-0218-01 Non-Interpretive 12-lead - 13 4. “FAX BUSY” message is displayed. Receiving phone line is being used. The M
OPTION INSERT 9650-0218-01 Non-Interpretive 12-lead - 14 The M Series with 12-Lead option may include a modem for transmitting 12-lead ECG informatio
Non-Interpretive 12-lead 9650-0218-01 Non-Interpretive 12-lead - 15 • The modem in the M Series is compatible with certain AMPS (analog) or dual-m
OPTION INSERT 9650-0218-01 Non-Interpretive 12-lead - 16 CAUTION • Data transmission via cellular phones can be less reliable than transmission via
Non-Interpretive 12-lead 9650-0218-01 Non-Interpretive 12-lead - 17 Interference WARNING Changes or modifications to this unit not expressly approv
This issue date or revision level for this operation guide is shown on the front cover. ZOLL and M Series are trademarks of ZOLL Medic
OPTION INSERT 9650-0218-01 Non-Interpretive 12-lead - 18 Shielded Cables The use of any cable other than the shielded type will allow your system t
9650-0218-01 Non-Interpretive 12-lead - 1 Rev. B ECG leads are a defibrillation protected Type CF patient connection. Product Description ZOLL M
OPTION INSERT 9650-0218-01 Non-Interpretive 12-lead - 2 WARNINGS • Before use, carefully read the M Series Operator’s Guide and these operating instr
Non-Interpretive 12-lead 9650-0218-01 Non-Interpretive 12-lead - 3 Electrode Placement Place electrodes on the patient. All electrodes must be con
OPTION INSERT 9650-0218-01 Non-Interpretive 12-lead - 4 12-LEAD ACQUISITION • Attach the electrodes to the patient (Refer to the “Electrode Placement
Non-Interpretive 12-lead 9650-0218-01 Non-Interpretive 12-lead - 5 If the M Series unit has been configured to automatically transmit 12-lead ECG d
OPTION INSERT 9650-0218-01 Non-Interpretive 12-lead - 6 0.05– 40 4x3 channel printout (0.05 – 40Hz 4x3): When programmed to this setting, all 4x3 12-
Non-Interpretive 12-lead 9650-0218-01 Non-Interpretive 12-lead - 7 Patient Identification Number (ID#) Pressing the ID # softkey causes the “Name a