12-Lead ECG Monitoring 9650-0215-01 Rev. H

Page 2

OPTION INSERT 9650-0215-01 12-lead - 8 12-Lead Reports Immediately after acquisition, the 12-lead data prints in the following order: 1. 12-Lead EC

Page 3 - 12-LEAD ECG MONITORING

12-lead 9650-0215-01 12-lead - 9 4x3 Median Complexes This format prints a single median beat for each of the twelve (12) leads. The median beat

Page 4 - WARNINGS

OPTION INSERT 9650-0215-01 12-lead - 10 Patient Information DEVICE ID: MEDIC_1RECORDED: 11:33:01 15 DEC 99PATIENT NAME : JOE SMITHPATIENT ID: 19

Page 5 - Electrode Placement

12-lead 9650-0215-01 12-lead - 11 12SL Wave Measurement After the P, QRS, and T complexes have been demarcated, the waves for each complex are i

Page 6 - Acquire Softkey

OPTION INSERT 9650-0215-01 12-lead - 12 Daily Operational Verification 1. Connect the V-lead cable to the 12-lead cable. 2. Connect the lead w

Page 7 - Settings Softkey

12-lead 9650-0215-01 12-lead - 13 Troubleshooting The troubleshooting guide is intended to help you identify and correct problems that arise duri

Page 8 - PT Info Softkey

OPTION INSERT 9650-0215-01 12-lead - 14 3. Unsuccessful transmission to fax (e.g., transmission stops before completion, signal disappears, etc.) 1.

Page 9

12-lead 9650-0215-01 12-lead - 15 8. “FAX INTERRUPTED” or “MUSE INTERRUPTED” message is displayed. 1. Power cycle M Se

Page 10 - 12-Lead Reports

OPTION INSERT 9650-0215-01 12-lead - 16 APPENDIX A MODEM AND PHONE SETUP The M Series with 12-Lead option may include a modem for transmitting 12-l

Page 11 - 12-lead

12-lead 9650-0215-01 12-lead - 17 requires a specific modem- to-cell phone interconnect cable. Cables may be purchased through The Supply Net, I

Page 12 - Measurements Matrix

This issue date or revision level for this operation guide is shown on the front cover. ZOLL and M Series are trademarks of ZOLL Medic

Page 13 - WARNING

OPTION INSERT 9650-0215-01 12-lead - 18 CAUTION • Data transmission via cellular phones can be less reliable than transmission via landline connec

Page 14 - OPTION INSERT

12-lead 9650-0215-01 12-lead - 19 Interference WARNING Changes or modifications to this unit not expressly approved by the party responsible for

Page 15 - Troubleshooting

OPTION INSERT 9650-0215-01 12-lead - 20 Canadian Requirements The Industry Canada, formerly the Canadian Department of Communications, label identi

Page 16

12-lead 9650-0215-01 12-lead - 21 APPENDIX B 12-LEAD ECG DATA TRANSMISSION TO CATALYST™ MUSE® SYSTEM The ZOLL M Series with 12-Lead and Catalyst

Page 17

9650-0215-01 12-lead - 1 Rev. H 12-LEAD ECG MONITORING ECG leads are a defibrillation protected Type CF patient connection. General Information

Page 18 - MODEM AND PHONE SETUP

OPTION INSERT 9650-0215-01 12-lead - 2 WARNINGS • Before use, carefully read the M Series Operator’s Guide and these operating instructions. • The

Page 19 - Configuring Phone Numbers

12-lead 9650-0215-01 12-lead - 3 Electrode Placement Location AHA Labels IEC LabelsRight Arm RA (white) R (red)Left Arm LA (black) L (yellow)Righ

Page 20 - CAUTION

OPTION INSERT 9650-0215-01 12-lead - 4 12-LEAD ACQUISITION • Attach the electrodes to the patient (Refer to the “Electrode Placement” Section). •

Page 21 - Shielded Cables

12-lead 9650-0215-01 12-lead - 5 If the M Series unit has been configured to automatically transmit 12-lead ECG data records to a fax machine or

Page 22 - Canadian Requirements

OPTION INSERT 9650-0215-01 12-lead - 6 not alter the bandwidth of ECG data passed to the 12SL program. It only affects the printed waveforms. Select

Page 23 - MUSE® SYSTEM

12-lead 9650-0215-01 12-lead - 7 Patient Identification Number (ID#) Pressing the ID # softkey causes the “Name and ID#” screen to appear. “Name”