12-Lead ECG Monitoring 9650-0215-01 Rev. H
OPTION INSERT 9650-0215-01 12-lead - 8 12-Lead Reports Immediately after acquisition, the 12-lead data prints in the following order: 1. 12-Lead EC
12-lead 9650-0215-01 12-lead - 9 4x3 Median Complexes This format prints a single median beat for each of the twelve (12) leads. The median beat
OPTION INSERT 9650-0215-01 12-lead - 10 Patient Information DEVICE ID: MEDIC_1RECORDED: 11:33:01 15 DEC 99PATIENT NAME : JOE SMITHPATIENT ID: 19
12-lead 9650-0215-01 12-lead - 11 12SL Wave Measurement After the P, QRS, and T complexes have been demarcated, the waves for each complex are i
OPTION INSERT 9650-0215-01 12-lead - 12 Daily Operational Verification 1. Connect the V-lead cable to the 12-lead cable. 2. Connect the lead w
12-lead 9650-0215-01 12-lead - 13 Troubleshooting The troubleshooting guide is intended to help you identify and correct problems that arise duri
OPTION INSERT 9650-0215-01 12-lead - 14 3. Unsuccessful transmission to fax (e.g., transmission stops before completion, signal disappears, etc.) 1.
12-lead 9650-0215-01 12-lead - 15 8. “FAX INTERRUPTED” or “MUSE INTERRUPTED” message is displayed. 1. Power cycle M Se
OPTION INSERT 9650-0215-01 12-lead - 16 APPENDIX A MODEM AND PHONE SETUP The M Series with 12-Lead option may include a modem for transmitting 12-l
12-lead 9650-0215-01 12-lead - 17 requires a specific modem- to-cell phone interconnect cable. Cables may be purchased through The Supply Net, I
This issue date or revision level for this operation guide is shown on the front cover. ZOLL and M Series are trademarks of ZOLL Medic
OPTION INSERT 9650-0215-01 12-lead - 18 CAUTION • Data transmission via cellular phones can be less reliable than transmission via landline connec
12-lead 9650-0215-01 12-lead - 19 Interference WARNING Changes or modifications to this unit not expressly approved by the party responsible for
OPTION INSERT 9650-0215-01 12-lead - 20 Canadian Requirements The Industry Canada, formerly the Canadian Department of Communications, label identi
12-lead 9650-0215-01 12-lead - 21 APPENDIX B 12-LEAD ECG DATA TRANSMISSION TO CATALYST™ MUSE® SYSTEM The ZOLL M Series with 12-Lead and Catalyst
9650-0215-01 12-lead - 1 Rev. H 12-LEAD ECG MONITORING ECG leads are a defibrillation protected Type CF patient connection. General Information
OPTION INSERT 9650-0215-01 12-lead - 2 WARNINGS • Before use, carefully read the M Series Operator’s Guide and these operating instructions. • The
12-lead 9650-0215-01 12-lead - 3 Electrode Placement Location AHA Labels IEC LabelsRight Arm RA (white) R (red)Left Arm LA (black) L (yellow)Righ
OPTION INSERT 9650-0215-01 12-lead - 4 12-LEAD ACQUISITION • Attach the electrodes to the patient (Refer to the “Electrode Placement” Section). •
12-lead 9650-0215-01 12-lead - 5 If the M Series unit has been configured to automatically transmit 12-lead ECG data records to a fax machine or
OPTION INSERT 9650-0215-01 12-lead - 6 not alter the bandwidth of ECG data passed to the 12SL program. It only affects the printed waveforms. Select
12-lead 9650-0215-01 12-lead - 7 Patient Identification Number (ID#) Pressing the ID # softkey causes the “Name and ID#” screen to appear. “Name”